The SBC uses a MAMA detector which is a photon-counting device that provides a two-dimensional ultraviolet capability. The MAMA accumulates photons in the ACS data buffer as they are received, producing images with 1024x1024 pixels. For a more detailed description, see the discussion of SBC ACCUM mode in the
ACS Instrument Handbook.
SBC must be specified if spectral element
PR110L or
PR130L is specified (STScI software will adjust the HST pointing to compensate for the refraction).
Unlike WFC, SBC exposures with a dispersing spectral element (PR110L or
PR130L) will not have an additional image exposure scheduled automatically . If an image exposure is desired in combination with a prism exposure, it must be entered as a separate exposure specification. It is best to place it immediately before or after the prism exposure in the same visit (see the
ACS Instrument Handbook for more details).
A direct imaging exposure is required for the prism wavelength calibration. Note that its bright-object constraints will be more stringent than those of the prism exposure, so the least sensitive filter feasible is recommended.
Note that Time_per_Exposure for individual SBC exposures must be an integer multiple of 0.1 second and must be in the range of 0.1 to 3600 sec.