Space Telescope Science Institute   13.8 Mode = ACQ/PEAKD Config = COS/NUV  13.10 Mode = ACCUM Config = COS/FUV

OBSERVING MODES


13.9 Mode = TIME-TAG
Config = COS/FUV


In TIME-TAG Mode, the COS FUV detector produces an event stream with a time resolution of 32 milliseconds. The X and Y pixel coordinates of each photon event are stored in a 32-bit word in data buffer memory. At the start of an exposure and after every subsequent 32-millisecond period which contains photon events, a 32-bit time-of-day word is written to the data memory. The TIME-TAG is the recommended data-taking mode unless the target's brightness leads to a total count rate in excess of that supported by this mode.

13.9.1 Aperture or FOV

The following apertures are allowed:

13.9.2 Spectral Element

Enter a spectral element from Table 13.3 for the COS/FUV configuration.

13.9.3 Wavelength

Enter the value of the central wavelength in Angstroms. Table 13.3 gives the allowed values of the central wavelength for each grating.

13.9.4 Optional Parameters

SEGMENT
= BOTH (default except for G140L, =1105 Å)
= A (default for G140L, =1105 Å), B

Specifies which segment of the FUV detector to use for an observation. A value of BOTH will activate both segments. If A is selected, only segment A of the detector will be activated for photon detection, and the spectrum will contain data from only that half of the detector. If B is selected, only segment B of the detector will be activated and used to generate data. The use of only a single segment may be warranted with sources that are too bright to observe safely over the entire detector, or when a source has a much higher expected count rate in one segment than the other and the important science is in the segment with the low count rate. See the COS Instrument Handbook for more information.

If you specify grating G140L with the 1105 Angstrom wavelength setting, SEGMENT defaults to A. For all other grating, central wavelength settings, BOTH is the default.

BUFFER-TIME
= 80 or greater (integer seconds)

Specifies the estimated minimum time in which 2.35 x 106 photon events (half of the COS data buffer capacity) will be accumulated during a TIME-TAG exposure. BUFFER-TIME is a required parameter if the target is not WAVE. If the target is WAVE, then BUFFER-TIME may not be specified .

If the predicted total number of events from a TIME-TAG exposure exceeds the COS data buffer capacity of 4.7 x 106 photon events, data must be transferred to the HST onboard science recorder during the exposure. Transfers of data from the COS buffer during an exposure will be made in 9-MByte blocks (half the buffer capacity). The value of BUFFER-TIME should be the half-buffer capacity (2.35 x 106 counts) divided by the estimated maximum average count rate in photons per second.

Note that BUFFER-TIME should include expected counts from the detector dark current and stim pulses (see the COS Instrument Handbook) as well as the detected photon events, factoring in the instrument quantum efficiency. A conservative value of BUFFER-TIME is recommended (err slightly on the low side) to avoid data loss.

The absolute minimum BUFFER-TIME of 80 seconds corresponds to a maximum average count rate of ~30,000 counts/sec over the entire detector, which is the maximum rate at which the flight software is capable of processing counts. The COS Exposure Time Calculator (ETC) can be used to estimate a suitable value for BUFFER-TIME. See the COS Instrument Handbook for more information.

EXTENDED
= NO (default), YES

Indicates whether the target is an extended source. This parameter is used only for data reduction.

FP-POS
= AUTO, 1, 2, 3 (default), 4

Indicates whether to take the exposure at a single offset, at multiple offsets, or at no offset from the nominal central wavelength. Obtaining exposures at small wavelength offsets from the specified central wavelength aids in the correction of the fixed-pattern defects of the detector. FP-POS=AUTO indicates that the specified exposure time will be divided evenly among four subexposures, and each subexposure will be obtained at a different predetermined offset from the specified central wavelength.

The values FP-POS=1, 2, 3, 4 will result in the exposure being taken at an offset from the specified central wavelength determined by the value. The default value, or if FP-POS is not specified on the exposure, will result in the exposure being obtained at the nominal central wavelength (i.e., FP-POS=3) offset. The exposure will be for the specified exposure duration. See the COS Instrument Handbook for specifics of the offsets and for guidance on how to use this capability.

Note for internal targets: FP-POS is not allowed for internal targets except Target=WAVE. Allowed values for exposures with Target=WAVE are FP-POS=1, 2, 3 (or not specified), or 4; FP-POS=AUTO is not allowed.

FLASH
= YES (default), NO

Indicates whether or not to "flash" the wavelength calibration lamp during exposures. These flashes are needed to compensate for the effect of post-move drift of the Optic Select Mechanisms. The default behavior will be that the wavecal lamp is turned on briefly at the start of an externally targeted exposure, and at intervals later in the exposure. The grating-dependent "flash" durations and the time-since-move-dependent flash intervals will be defined by the STScI. Specifying the value NO will disable automatic flashing for the current exposure.

When flashing is enabled, the exposure time must be at least as long as a single flash. Flash durations, as functions of grating and central wavelength, are given in Table 13.5.

When aperture BOA is selected, FLASH may not be specified and defaults to NO.

13.9.5 Number of Iterations

Number_Of_Iterations must be 1 in this Mode.

13.9.6 Time Per Exposure

Enter the total time of data collection as Time_Per_Exposure. Time_Per_Exposure must be an integral multiple of 0.1 seconds (if it is not, its value will be rounded down to the next lower integral multiple of 0.1 sec, or set to 0.1 seconds if a smaller value is specified), and may range from 0.1 to 6500 seconds. If the exposure specifies FP-POS=AUTO, then the valid range is 0.4 to 26000.0 seconds. Values much larger than 3000 seconds (12000 seconds if FP-POS=AUTO) are normally appropriate only for visits with the CVZ special requirement.

For Target=WAVE, enter DEF for Time_Per_Exposure.

If BUFFER-TIME < 110 seconds, photon events may be generated faster than data can be transferred out of the buffer during the exposure. In this case, Time_Per_Exposure should be less than or equal to 2 * BUFFER-TIME so that the exposure can complete before data transfer is necessary. A BUFFER-TIME of 110 seconds corresponds to an average count rate of ~21,000 counts/sec.

Note that TIME-TAG exposures have the potential to rapidly use up the HST onboard storage capacity. Caution is advised on any exposure with an exposure time greater than 25 * BUFFER-TIME, which corresponds to ~6 x 107 counts, or about 2 GBits (close to 20% of the solid-state recorder capacity).

13.9.7 Special Requirements

The special requirement RT ANALYSIS is not permitted on TIME-TAG exposures.


 13.8 Mode = ACQ/PEAKD Config = COS/NUV  13.10 Mode = ACCUM Config = COS/FUV
Space Telescope Science Institute
http://www.stsci.edu
Voice: (410) 338-1082
help@stsci.edu