ACQ/PEAKXD Mode invokes a procedure in the COS onboard flight software that places the target near an optimal position within the selected aperture in the cross-dispersion direction on the detector. This procedure consists of several steps.
ACQ/PEAKXD Mode should normally be preceded by an ACQ/SEARCH to ensure that the target is positioned in the aperture. Details on the target acquisition sequence and limiting magnitudes for the PSA and BOA can be found in the COS Instrument Handbook.
Subarrays are used during target acquisition to avoid geocoronal airglow lines, which could bias the target acquisition calculations, as well as portions of the detector which are not illuminated by the input spectrum. The subarray dimensions depend on the selected grating and central wavelength; see the COS Instrument Handbook for details.
The aperture used for the ACQ/PEAKXD Mode exposure need not be the same as the aperture used for the subsequent TIME-TAG or ACCUM science exposures. The following apertures are allowed:
As with spectroscopic observations, use of the BOA is recommended on relatively bright targets to attenuate the flux and allow the acquisition to proceed without triggering bright object violations.
Enter a spectral element from Table 13.3 for the COS/FUV configuration. This will generally be the same as the one used for the subsequent TIME-TAG or ACCUM science observation.
Enter the value of the central wavelength in Angstroms. Table 13.3 gives the allowed values of the central wavelength for each grating.
=1105 Å), B
This is the same as in ACQ/SEARCH Mode; see Section 13.2.4.
The Number_Of_Iterations must be 1.
Enter the total time of data collection as Time_Per_Exposure. Time_Per_Exposure must be an integral multiple of 0.1 seconds. If it is not, its value will be rounded down to the next lower integral multiple of 0.1 sec, or set to 0.1 seconds if a smaller value is specified.
The procedures to determine the exposure time for the ACQ/PEAKXD Mode exposure are given in the COS Instrument Handbook.
The special requirement POSition TARGet <X-value>,<Y-value> is not permitted on ACQ/PEAKXD exposures.
|
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |