FOS Instrument Science Report # 118 "SMOV Report III: FOS Baseline Sensitivity" by Charles (Tony) Keyes, Anne Kinney, Anuradha Koratkar, and Cynthia Taylor Abstract excerpted from the report: Following Servicing Mission (SM) but before COSTAR delployment, spectrophotometric standard BD+28 4211 was observed with the 4.3 aperture and several representative detector/dispenser combinations. The observations were compared with Cycle 3 measures obtained with identical target acquisition and observational setups. Measured FOS/BLUE and FOS/RED count rates were higher in the post-SM observations by approximately 7% and 6% respectively. The increase in post-SM instrumental sensitivity is consistent with predictions based upon HST secondary mirror re-positioning in the post-SM time period. FOS instrumental sensitivity is unaffected by any SM-related activities. For more information, please contact the above authors.