The average full well depths for the ACS CCDs are given in Table 3.1 as 84,700 e
− for the WFC and 155,000 e
− for the HRC, but the pixel-to-pixel values vary by ~10% and ~18% across the fields of view of the WFC and the HRC, respectively. When the CCD is over-exposed, pixels will saturate and excess charge will flow into adjacent pixels along the column. This condition is known as “blooming” or “bleeding.” Extreme overexposure does not cause any long-term damage to the CCDs, so there are no bright object limits for the ACS CCDs. When using ADC gains of 2 for the WFC and 4 for the HRC, the linearity of the detectors deviates by less than 1% up to the full well depths. On-orbit tests using aperture photometry show that this linearity is maintained when summing over pixels that surround an area affected by charge bleeding, even when the central pixel is saturated by 10 times its full well depth (see
ACS ISR 2004-01 for details).
Prior to failure of the original CEB, the WFC read noise had been generally constant with time. After a transit through the South Atlantic Anomaly (SAA) on 29 June 2003, the read noise of amplifier A changed from ~4.9 to ~5.9 e− rms. Although the telemetry did not show any anomaly in any WFC components, it is likely that the sudden increase was caused by radiation damage. Amplifier A was the only amplifier that showed this anomaly. The amplitude of the variation (~1 e
−) was the same for ADC gains 1 and 2. After the following anneal (see
Section 4.3.5), the read noise dropped to ~5.5 e
− and remained constant for 27 days. After the following two anneal cycles, the read noise stabilized at ~5.6 e
−.
Figure 4.4 shows the read noise in the image area for amplifier A during the instability period. The read noise of the other amplifiers was very stable between the installation of ACS in March 2002 and the failure of ACS in January 2007.
Prior to January 2007, the read noise of the HRC was monitored using only the default amplifier C and the default gain of 2 e−/DN. No variations were observed with time. The read noise measured in the physical overscan and image areas were consistent with the pre-flight values of 4.74 e
− (see
Table 4.2 and
Table 4.3).
All ACS CCDs are buried channel devices which have a shallow n-type layer implanted below the surface to store and transfer the collected signal charge away from the traps associated with the Si-SiO2 interface. Moreover, ACS CCDs are operated in Multi-pinned Phases (MPP) mode so that the silicon surface is inverted and the surface dark current is suppressed. ACS CCDs therefore have very low dark current. The WFC CCDs are operated in MPP mode only during integration, so the total dark current figure for WFC includes a small component of surface dark current accumulated during the readout time.
Like all CCDs operated in a Low Earth Orbit (LEO) radiation environment, the ACS CCDs are subject to radiation damage by energetic particles trapped in the radiation belts. Ionization damage and displacement damage are two types of damage caused by protons in silicon. The MPP mode is very effective in mitigating the damage due to ionization such as the generation of surface dark current due to the creation of trapping states in the Si-SiO2 interface. Although only a minor fraction of the total energy is lost by a proton via non-ionizing energy loss, the displacement damage can cause significant performance degradation in CCDs by decreasing the charge transfer efficiency (CTE), increasing the average dark current, and introducing pixels with very high dark current (hot pixels). Displacement damage to the silicon lattice occurs mostly due to the interaction between low energy protons and silicon atoms. The generation of phosphorous-vacancy centers introduces an extra level of energy between the conduction band and the valence band of the silicon. New energetic levels in the silicon bandgap have the direct effect of increasing the dark current as a result of carrier generation in the bulk depletion region of the pixel. As a consequence, the dark current of CCDs operated in a radiative environment is predicted to increase with time.
At the beginning of the side-2 operation in July 2006 the temperature set point of the WFC was lowered from -77 °C to -81
°C (documented in
ACS TIR 2006-02). Following SM4, we measure a dark current of 22 - 25 e
−/pixel/hour among the four amplifiers. We have reverted to 12-hour anneals.
We have chosen to classify the field-enhanced pixels into two categories: warm and hot pixels. The definition of “warm” and “hot” pixel is somewhat arbitrary. We have chosen a limit of 0.08 e–/pixel/seconds as a threshold above which we consider a pixel to be “hot”. We identify “warm” pixels as those which exceed by about 5 σ (was 0.02 e
–/pixel/second but after SM4 it is 0.04 e
–/pixel/second) the normal distribution of the pixels in a dark frame up to the threshold of the hot pixels (See
Figure 4.6) for a typical dark rate pixel distribution.
Most of these new hot pixels are transient. Like others CCDs on HST, the ACS devices undergo a monthly annealing process. (The CCDs and the thermal electric coolers are turned off and the heaters are turned on to warm the CCDs to ~19
°C.) Although the annealing mechanism at such low temperatures is not yet understood, after this “thermal cycle” the population of hot pixels is greatly reduced (see
Figure 4.5). The anneal rate depends on the dark current rate; very hot pixels are annealed more easily than warmer pixels. For pixels classified as “hot” (those with dark rate > 0.08 e
–/pix/sec) the anneal rate is ~82% for WFC and ~86% for HRC.
For example, a simple ACS-WFC-DITHER-LINE pattern has been developed that shifts the image by 2 pixels in X and 2 pixels in Y along the direction that minimizes the effects of scale variation across the detector. The specific parameter values for this pattern are given on the ACS dithering Web page at:
The CTE numbers for the ACS CCDs at the time of installation are given in Table 4.6. While the numbers look impressive, remember that reading out the WFC CCD requires 2048 parallel and 2048 serial transfers, so that almost 2% of the charge from a pixel in the corner opposite the readout amplifier was lost.
Like other CCD cameras aboard HST, ACS has suffered degraded CTE due to radiation damage since its installation in 2002. Since 2003, specific calibration programs aimed at characterizing the effects of the CTE on stellar photometry have been performed. Results from the internal CTE calibration programs showed that CTE appears to decline linearly with time, and that CTE losses are stronger for lower signal levels (Mutchler & Sirianni,
ACS ISR 2005-03). However, in order to derive an accurate correction formula for photometry, a number of observations at different epochs and for different levels of sky background and stellar fluxes have to be accumulated. The results of the observations taken through November 2011 are described in several ISRs (
ACS ISR 2009-01,
ACS ISR 2011-01,
ACS ISR 2012-05) and are summarized here. For WFC, significant photometric losses are apparent for stars undergoing numerous parallel transfers (y-direction). The losses as of 2011 are between 5-10% for typical observing parameters, rising to ~50% in worst cases (faint stars, low background). The size of the photometric loss appears to have a strong power-law dependence on the stellar flux, as seen in other CCDs flown on
HST.
It may also be useful, in some specific cases, to choose the filter in order to obtain a higher background (e.g. F606W). CTE degradation also has an impact on astrometry (see ACS ISR 2007-04). Therefore, for astrometric programs of relatively bright objects, the use of post-flash may be considered. No losses are apparent for WFC due to serial transfers (x-direction). Correction formulae are presented in
ACS ISR 2012-05 to correct photometric losses as a function of a source’s position, flux, background, and time. Users are encouraged to check the
ACS Web page for updates. Further details on CTE corrections of very bright stars on subarrays can be found in the
ACS ISR 2011-01 (Bohlin & Anderson, 2011). The study of the CTE evolution is presented in
ACS ISR 2012-03 (Ubeda & Anderson, 2012).
In the optical, each photon generates a single electron. However, in the near-UV, shortward of ~3200 Å there is a finite probability of creating more than one electron per UV photon (see Christensen, O., J. App. Phys.
47,689, 1976). At room temperature the theoretical quantum yield (i.e., the number of electrons generated for a photon of energy E > 3.5eV (
λ~3500 Å)), is Ne = E(eV)/3.65. The HRC CCDs quantum efficiency curve has not been corrected for this effect. The interested reader may wish to see the
STIS Instrument Handbook for details on the signal-to-noise treatment for the STIS CCDs.