| Space Telescope Science Institute |
| COS Instrument Handbook |
| help@stsci.edu |
Table 1.1: COS Instrument Characteristics
Table 1.2: COS Detector Characteristics
Cs2Te (semi-transparent) MgF2 (re-entrant) 85 × 10 mm (two) 25.6 × 25.6 mm 16384 × 1024 (two) 1024 × 1024 1024 × 1024 6 × 10 pix 3 × 3 pix Typical dark-count rate (away from SAA)1 Detector global count rate limit2 ~15,000 cnt s–1 per segment 7.4 μsec
Count rates higher than these limits will trigger a detector shut-down. Bright-object screening limits are lower. See Chapter 10.
These values are not applicable to the COS FUV G130M blue cenwaves, 1055 and 1096. We expect the local count rate limits for those modes to be lower by a factor 2-3. However, the calibration data allowing the derivation of the local count rate limits for the 1055 and 1096 cenwaves are still being worked on. Users should check the COS Web site for updates on those values.
Table 1.3: COS Calibration Accuracies
15 km s–1 15 km s–1 150 km s–1 175 km s–1 15 km s–1 15 km s–1 Relative photometry (same object at a different time) Flat field quality (spectral S/N) per resel, using standard techniques 35:11 Flat field quality (spectral S/N) per resel, goal
Table 1.4: Useful Figures and Tables
HST focal plane and COS aperture Example of 3 × 3 spiral search pattern Overheads and observing parameters