Table 5.2 summarizes Transfer mode sources of error and associated calibrations. Each entry is described in the subsections below.
Targets fainter than V ~ 14.5 are increasingly affected by dark counts which reduces the amplitude of the S-Curve (since these contributions are not coherent with the light from the star. see
Chapter 7). The instrumental dark counts have been measured as part of the STScI FGS calibration program (and are listed in
Table 2.1). These values are needed for the analysis of data from Transfer mode observations of stars with V>14.
Drift is defined in Chapter 7, and its application to Position mode observations has been discussed earlier in this chapter. Drift is also apparent in Transfer mode observations but its removal from the raw data is straightforward; the cross-correlation of S-Curves, prior to binning and co-adding, automatically accounts for drift. Each individual S-Curve is shifted so that the particular feature of the S-Curve used for the cross correlation coincides with that of the fiducial S-Curve. However, the reliability of implicitly removing the drift is only as good as the accuracy of the cross correlation procedure, which becomes photon noise dominated for stars fainter than V=15 (i.e., cross correlation of scans is not reliable for V>15, hence drift can not be removed from such observations).
Figure 5.5 illustrates FGS3’s persistent variability and FGS1r’s stability over comparable timescales.
Figure 5.5a shows FGS3’s inherent variability in the S-Curves of the same point source over a 102-day span in 1997. Note this intrinsic variability in the instrument is indistinguishable from its interferometric response to a 15 mas binary with Δm = 0.5 mag (when compared with a point-source, as shown in
Figure 5.5c). In contrast, the stability of FGS1r (
Figure 5.5b) easily permits detection of the 15 mas binary system (
Figure 5.5d) when compared to a point-source. We note the “difference” seen in
Figure 5.5d are due to
systematic changes in the observed interferogram (i.e., the object is resolved), not random effects.
The morphology and amplitude of the S-Curve is sensitive to the spectral color of the source. The point-source reference S-Curve used for comparison to the observed fringes of a science target should match the color of the target to within δ(B - V) = 0.3, especially when analyzing observations of close binaries (separations < ~40 mas). The point source standards observed by FGS1r as part of the yearly observatory calibration plans are chosen to meet the needs of the cycle’s GO science program. This includes consideration of both the science target colors as well as the filter (F583W or F5ND) to be used by the GOs. For example, the cycle 10 FGS1r reference star color library is given in the
Table 5.3.