FOC Linearity
If two or more photon events overlap during a given frame, the VPU detection
logic will only count one detected photon. This `undercounting', or non-linearity,
sets a hard limit on the maximum allowable photon rate for the
FOC. The linearity performance also depends on the image structure, with
extended sources and point sources having unique behaviors. Regardless
of the structure of the source, the size of the image determines the
basic linearity performance characteristics. Therefore, by choosing a
small format, the user can maximize the count rate that can be observed
without suffering from non-linearity. Complete details can be found
in either Section 6.2 of the FOC Instrument Handbook Version 5.0, or in one
of the Instrument Science Reports written on linearity.
Extended Source (Flat-Field) Behavior
The flatfield nonlinearity was measured on-orbit during OV and SV using
observations of the internal LED calibration lamps. The linearity curve
was measured for several different formats to verify the format dependence
of the saturation. The measured linearity relation for the 512 X 512 formats
of the pre-COSTAR f/96 and pre-COSTAR f/48 detectors was fit by the
relation:
Values of a for the most commonly-used formats are given as:
The values refer to dezoomed data in the case of formats that were
originally zoomed.
To ensure that non-linearity does not compromise
the science data, users are advised to ensure that the count rate is
kept below N(MAX), which is the count rate that would give 10%
nonlinearity, as given in the third column of the table. Correct and
quantifiable operation of the FOC at count rates exceeding N(MAX)
cannot be guaranteed.
Point-Source Behavior
When the illumination comes from a star, the FOC is able to count at a
much higher rate before saturation occurs. During OV and SV it became
clear that for the FOC pre-COSTAR f/96
camera when used in the 512 X 512 format, the maximum obtainable
count rate in the core of a star was approximately 3
counts/second/pixel. At count levels higher than this, the core of the star
turns into a dark 'hole', and a bright crescent appears to one side of
the core.
Comparison of PSF's taken with and without neutral density
filters indicate that there is no noticeable deviation from linear
behavior for core count rates up to 1 count/sec in the brightest
pixel, as seen in this figure:
The use of either SYNPHOT or FOCSIM to calculate the
count rate for the source can provide information on the expected linearity
for the image. The simple guideline for observers should be
keep the count rate in the central pixel below 1 count/sec.
By following this guideline, any photometry method chosen will give
results that are not biased by non-linearity effects.
|