STScI Logo
STScI Logo
HST
Banner
FAQStaff

NICMOS Coronagraphic Observing for Cycle 13 onwards

Note for coronagraphic observers.

Recommendations for ACQ exposure times can be found in Section 5.1.1 of the NICMOS Instrument Handbook. In particular,  Figure 5.4 summarizes the exposure time  recommendations for the  F160W filter acquisitions.

If the NICMOS (imaging) ETC is used to this purpose, users should take into account that:

  1. When acquiring objects bright enough so that allocation of exposure time is not an issue, observers should plan target acquisition exposures to obtain ~ 77,000 e- in the central pixel of the PSF. This should not be exceeded for risk of a TA failure. For fainter targets, more shallow exposures will result in inferior target/hole positioning due to poorer target image centroiding, but is robust down to levels of several thousand e-.(see Instrument Handbook)
  2. All ETC calculations for NIC2 are relative to the NIC2 focus and not at the coronagraphic focus.  The predicted count rate will need to be reduced by ~6% for the F160W filter at the coronagraphic focus. The above limit (i.e. 77000e-) already takes this correction into account.

Shortcuts

Instrument Handbook
Data Handbook
Handbooks Errata
ISRs
Advisories
Anomalies
NICMOS ETC
Dark Generator
Flat Generator
STScI Acronym List


Copyright  | Help  | Printable Page