Because of the effects of shading, and the possibility that the underlying dark current may vary with time since reset, the removal of dark current (for calibration purposes, we implicitly assume shading and amplifier glow is a part of the time variable “dark current”) from NICMOS data is more complicated than for other instruments. The most accurate way to remove the dark current from any observation is a measurement of the dark current with an identical integration time and at the same detector temperature. For
ACCUM observations we encourage observers to select exposure times that correspond to existing delta-times of the pre-defined
MULTIACCUM sequences (
Table 8.2). These available delta-times are also listed in
Table 8.3 below.
STScI is not planning to obtain dedicated ACCUM dark observations, but dark reference files for
ACCUM observations can be constructed from
MULTIACCUM darks with identical delta-times.