Table 13.1: Summary of Cycle 11 Calibration Goals
| Attribute |
Accuracy |
Limiting Factor/Notes |
| Detector dark current and shading |
<10 DN |
All MULTIACCUM sequences will have good calibrations. A small subset of the available ACCUM mode exposure times will have a direct calibration. |
| Flat fields |
1% broad-band 3% narrow-band |
Color and temperature dependence may limit flats in some cases. The low spatial frequency may only be reliable to 3%. |
| Photometry |
6% photometric zero point 2% relative over FOV 2% stability |
Photometric systems, Intrapixel effects, Filter leaks on red sources, Temperature stability |
| PSF and focus |
Maintained within 1 mm (except for NIC3 which will be ~ 4mm off nominal best focus.) |
Breathing and OTA desorption are approximately equal effects. Cold-mask shifting affects PSF shape. |
| Coronagraphic PSF |
0.013 arcsec accuracy positioning in the hole |
|
| GRISM wavelength calibration |
.005 µm |
Limited by centroiding of target for wavelength zeropoint determination. |
| GRISM photometric calibration |
~30% absolute and relative |
The figure is quoted without the NIC3 intrapixel sensitivity correction. With this correction, GRISM photometry will be better. Expected accuracy over central 80% of spectral range. Grism C flat field may not achieve this performance. |
| Astrometry |
0.5% plate scale 0.1" to FGS frame |
per observation, after geometric distortion correction |