|
Many of the actual characteristics of the NCS such as power consumption,
lowest possible operating temperature, or temperature stability, are
still uncertain and have to await on-orbit testing. The numbers stated here
should be regarded as estimates only.
The cryocooler, manufactured by Creare, Inc., is a reverse-Brayton cycle
turbine design. The compression and subsequent expansion of the Neon gas
results in a net cooling which can be used to remove heat from the NICMOS
dewar via the heat exchanger to the circulator loop (
see
Figure 2.5 in the Instrument Handbook).
The Creare design has several major advantages for application on HST. First,
the closed loop system operates at very high speed -- about 7000
revolution per second -- which limits the probability of mechanical
coupling to the HST structure, thus minimizing the risk of spacecraft
jitter. Second, the system is capable of providing large cooling power.
In order to maintain NICMOS at an operating temperature around
75-86 K, an estimated cooling power of 400 mW is needed. Since the
parasitic losses due to the flex lines and bayonet couplings (see below)
are rather large, delivering 400 mW of cooling power to NICMOS requires
about 8 W from the cooler. Finally, the Creare design is compact enough
to fit into the previously unused space between the NICMOS enclosure and
the HST aft end bulkhead.
The cryocooler dissipates between 250 and 400 W
of energy which needs to be removed from the aft shroud. This is achieved
via a Capillary Pumped Loop (CPL) that thermally connects the housing of
the compressor pump with an external radiator. The continuous heat flow
through the (passive) CPL lines is maintained by a set of heaters that are
controlled by the ESM micro controller. The heat from inside NICMOS is
transported to the cryocooler via a set of flex lines. The flex lines
connect to the inside of the dewar via bayonet fittings at an interface
plate outside of the dewar which is accessible to the astronauts during
SM3B.
After installation, this loop -- the circulator loop -- will be
filled with Neon gas from high pressure storage bottles. The Neon gas,
driven by a high speed circulator pump
and cooled via the heat exchanger to the cryocooler loop, circulates
through the cooling coil at the aft end of the NICMOS dewar,
thus cooling the entire instrument. The exact
temperature of the Neon gas in the circulator loop - and thus the
temperatures reached by the NICMOS detectors - will depend on the nature
and amount of the parasitic heat loads to the flex lines and the NICMOS
dewar. Current best estimates are 62 K at the heat exchanger, with about
400 mW of parasitic heat loads, this should result in detector
temperatures between 75 and 86 K.
The ESM will use the telemetry
readings of a number of temperature sensors inside the NICMOS dewar
in an active control law in order to regulate the cooling power of the
NCS, and thus provide stability of the operating temperature of the
NICMOS detectors. Because of the sensitive dependence of a number of
detector characteristics on temperature, the temperature
stability is a crucial requirement on NCS performance. The stability
specification of 0.1 K/orbit - combined with a setpoint repeatability
of 0.5 K - should allow stable calibration of NICMOS data. All
components of the NCS are combined into a common enclosure, the
NICMOS CryoCooler (NCC). The system has been
successfully tested in space during the HST Orbital Systems Test (HOST)
shuttle mission in fall 1998.
Remaining uncertainties about the NCS
performance stem from the natural lack of tests with the actual NICMOS
dewar. In ground tests, as well as during HOST, a NICMOS simulator was
used to mimic the expected parasitic heat loads from, and the mechanical
connections to the actual NICMOS. How closely this simulator resembles
the NICMOS dewar is somewhat uncertain, hence we expect a range of
possible operating temperatures for the revived NICMOS detectors.
|