Bias Jumps or Bands
DIAGNOSTIC:
It is currently believed that the bands are caused by a pull-down of the
amplifier bias when the amplifiers of another NICMOS camera are being
used at the same time. Since the readouts of parallel NICMOS exposures
are staggered by fractions of readout time (~1/3 of a readout time), then
bands will appear where the readout of another camera stops or starts.
The bands appear in approximately the same place on the detector, but are
not 100% spatially repeatable due to timing pattern differences. Also, the
polarity of the bias jumps is not always in the same sense or of the same
amplitude, making it somewhat non-repeatable.
CURE:
Since this feature is often (but not always) found in the final read of a
multiaccum sequence, it can usually be minimized by recalibrating the image,
eliminating the last read. If the loss in exposure time is important, then
running CALNICA up to the flatfielding step and subtracting
the column (or row, whichever is perpendicular to the fast read direction)
average of the image (after expanding this average to a 256x256 image) and
finishing the CALNICA processing can largely eliminate this feature. Note
that the noise caused by the increased bias will remain, however. In
this NIC3 example, the point source is (fortuitously) in a low bias
region, and peaks at over 28,000 cts/sec while the bias jumps between
-1, 11, and 23 cts/sec. Correction may not be necessary in this case.
Since Cycle 11, we have implemented a modified readout sequence for the three
NICMOS cameras which reduces the probability that a detector is being reset
while another is read. This procedure will be completely transparent to users
and should significantly reduce the electronic bands problem.
|
 |
|