Detector Count Rate Restrictions

 

The ETC uses the current count rate screening restrictions, which are taken from Section 11.5 of the COS Instrument Handbook. If the observation will exceed any of the count rate restrictions, a warning message will be given in the output page. The count rate restrictions for the detectors are as follows:

 

COS Count Rate Screening Limits

Detector

Source type 1

Type of limit

Limiting count rate 2

 

FUV

Predictable

Global

15,000 per segment

Local

40 per resel 3

Irregular

Global

6,000 per segment

Local

40 per resel 3

 

NUV

Predictable

Global

30,000 per stripe

Local

70 per pixel

Irregular

Global

12,000 per stripe

Local

70 per pixel

Imaging

Global

170,000 (over entire detector)

Local

70 per pixel

 

1"Predictable" means the brightness of the source can be reliably predicted for the time of observation to within 0.5 magnitude.

2Entries are counts per second.

3An FUV resel is 6 pixels wide by 10 high.

 

 

 

Detector Dark Background

The following table lists the dark count rate and read noise characteristics of the COS detectors as measured in ground tests.

Detector background count rates (per second) for COS.

Detector:

FUV XDL

NUV MAMA

Dark rate
(counts sec-1)

1.5 per cm2
7.2 x 10-7 per pixel
4.3 x 10-5 per resel

11 per cm2
2.1 x 10-4 per pixel
1.9 x 10-3 per resel

Read noise

0

0