You cannot use the entire target visibility time for actual science exposures, because of the required times for guide star acquisition, target acquisition, and SI overheads. The following subsections discuss the amounts of time that should be budgeted for these items; they are conservative approximations suitable for use in a Phase I proposal and may differ slightly from the numbers in the Instrument Handbooks.
A normal guide star acquisition, required in the first orbit of every visit, takes 6 minutes. At the beginning of subsequent orbits in a multi-orbit visit, the required guide star
re-acquisition takes 4 minutes. For CVZ observations guide star re-acquisitions are not required but, if an observation extends into SAA-impacted orbits (see
Section 2.2.2), then guide star re-acquisitions will be necessary for those orbits.
A target acquisition may be required after the guide star acquisition, depending on the SI used and pointing requirements. See
Section 5.2 for a basic overview of target acquisitions. Consult the
HST Instrument Handbooks to determine whether a target acquisition is required for your particular observations, and which acquisition type is most appropriate. Then use
Table 6.3 to determine the time that you need to budget for this.
The most common use of target acquisition is for COS and STIS spectroscopy. A number of target acquisition options are provided for each instrument. In some cases an additional set of peak-up observations can follow the initial instrument acquisition.
Most normal imaging observations with ACS, STIS and WFC3 do not require target acquisitions (assuming that the coordinates delivered by the observer in Phase II have sufficient accuracy of 1"-2"). As the COS aperture has a small 2.5” diameter field of view, a target acquisition is recommended for COS imaging unless the coordinates supplied by the observer are accurate to 0.4” or better. For COS imaging, the same acquisition strategies are available as for spectroscopy.
For
coronagraphic imaging with STIS you will need to perform a target acquisition to place the target behind the coronagraphic hole or feature. For STIS, the same ACQ and ACQ/PEAK strategies are available as for spectroscopy.
FGS observations use a so-called spiral search location sequence for target acquisitions. This is part of a science observation, and the time required for the acquisition is considered to be part of the overhead associated with the science observation (see
Table 6.6).
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Typical precision acquisition in dispersed light
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Used for STIS spectroscopic observations in apertures <= 0.1" in size, or for any STIS spectroscopic observation that requires the highest possible absolute precision in the zeropoint of the wavelength scale.This type of target acquisition always follows an ACQ. For faint targets (V > 20), add 4 times the acquisition exposure time determined by the Target Acquisition ETC.
|
|
|
|
|
Used for real time interactions with the telescope in very exceptional circumstances.
|
Generally, a target acquisition does not need to be repeated for separate orbits of a multi-orbit visit. However, we recommend that observers planning multi-orbit observations in 0.1" or smaller STIS slits insert a target peakup maneuver every 4 orbits (see
Section 3.2.1).
A target acquisition, if necessary, usually should be inserted in each visit. However, programs with multiple visits to the same target within a six-week period (start to finish) may be able to use the
Reuse Target Offset function (see
Section 5.2.2). If reuse target offset is appropriate for your program, then you should include the full target acquisition sequence only in the initial visit; the subsequent visits will not need a full target acquisition. However, they will require a
Small Angle Maneuver (SAM) (see
Section 6.4.4) for the offset maneuver, and they usually require the final peakup stage used in the original acquisition. Please contact the
STScI Help Desk (see
Section 1.4) if you feel your program can benefit from this capability.
There are a variety of instrument overheads associated with science exposures. Tables 6.4 to
6.9 summarize for each instrument how much time you need to budget for these overheads, depending on the observing strategy.
For several years, many observers have been using dithering, or small spatial displacements, to allow for better removal of detector defects and the reconstruction of sub-pixel resolution. In general,
undithered observations with the ACS CCD and WFC3 detectors will not be approved without strong justification that such is
required for the scientific objectives (see
Section 5.2.)
ACS exposure overheads are listed in
Table 6.4.The overhead per exposure is shorter if the exposure is the same as the previous exposure (i.e. the exposures use the same aperture and spectral element, but not necessarily the same exposure times). If you are unsure whether the shorter overhead time is appropriate, then use the longer overhead time (to avoid a possible orbit allocation shortfall later).
Note that if AUTOIMAGE=NO is invoked and a different direct image is specified for the WFC spectroscopic calibration, and in all cases for the SBC calibration (for which there is no AUTOIMAGE due to the safety issue), these direct images must be included explicitly in the Observing Summary and the observing time (orbit) request of the Phase I proposal.
An ACQ/SEARCH will require about 7 minutes of overhead, while an ACQ/Image requires 3 minutes. The combination of ACQ/PEAKXD and ACQ/PEAKD will also require 7 minutes. The first science exposure in a visit requires 5 minutes, while subsequent identical exposures incur 2 minutes of overhead. An additional 1 minute is needed for each instrument change between exposures, except that incrementing the FP-POS at the same CENWAVE setting only requires 3 seconds if the ordering of the FP-POS positions is done correctly. COS exposure overhead are listed in
Table 6.5
Table 6.5: COS Exposure
Overheads
FGS overheads are listed in Tables 6.6 and
6.7. The total TRANS mode overhead consists of an acquisition overhead plus an overhead per scan. Hence, the total overhead depends on the number of scans obtained during a target visibility period. In
Table 6.8 we list the recommended number of scans as a function of target magnitude. The recommended
exposure time is 40 seconds per scan (excluding overheads).
|
|
|
|
|
|
|
If target magnitude V < 14
|
|
|
|
If target magnitude 14 < V < 15
|
|
|
|
If target magnitude 15 < V < 16
|
|
|
|
If target magnitude 16 < V < 16.5
|
|
|
|
If target magnitude V > 16.5
|
|
|
|
Target acquisition (independent of target magnitude)
|
|
|
|
Overhead per scan (independent of target magnitude)
|
STIS overheads are listed in Table 6.9. The overhead per exposure is shorter if the exposure is the same as the previous exposure (‘no change’); this means that the exposures use the same aperture, grating and central wavelength, but the exposure times need not be the same. If you are unsure whether the shorter overhead time is appropriate, then use the longer overhead time.
WFC3 exposure overheads are listed in
Table 6.10.
Table 6.10: WFC3 Instrument Overhead Times
A “Reuse Target Offset” visit (see Section 5.2.2 and
Section 6.4.2) will require a SAM to be scheduled at the start of the first orbit. To allow for the offset adjustment, the SAM should be assumed to have a duration of 30 seconds.
Patterns described in
Section 5.4 perform a series of SAMs. The timing and subsequent overheads depend on the size of the pattern. However, a simple estimate for the overhead time associated with a pattern is obtained by multiplying the number of points minus 1 (one) times the overhead time for a single SAM (see
Table 6.11) whose size matches the pattern spacing.