|
-
|
G230L (MAMA) or G230LB (CCD) at 1600–3100 Å.
|
|
-
|
G230M (MAMA) or G230MB (CCD) at 1650–3100 Å.
|
Supported for use with the first-order gratings are long slits of widths 0.05, 0.1, 0.2, 0.5 and 2.0 arcseconds (in the dispersion direction), and lengths of 52 arcseconds (as projected on the CCD detector) or 25 arcseconds (as projected on the MAMA detectors) for the MAMA low-resolution, first-order gratings (G230L and
G140L) and 28 arcseconds for the MAMA intermediate-resolution, first-order gratings (
G230M and
G140M).
1 Note that the 0.1 arcsecond width matches the 2 pixel resolution of the CCD, while the 0.05 arcsecond width does so for the MAMAs, providing maximum spectral resolution. The 0.2 arcsecond-wide slit is the general utility slit used most often; it provides a good compromise between resolution and throughput. Programs requiring accurate absorption line measurements in continuum sources should always use slits of widths
≤0.2 arcsecond, since for larger apertures the spectral purity is significantly degraded by the telescope plus instrumental point spread function (PSF); see
Section 13.7. Finally, we expect the wider 0.5 and 2.0 arcsecond slits to be used predominantly in photon-starved UV observations of extended sources, but provide them for use in the optical as well to assure that line-ratio studies with coverage from the UV to the optical can sample the same physical region on the sky. Additionally, they are the most photometric slits as their throughput is least affected by centering and telescope breathing. Of course, observations of
extended sources with wide slits will have correspondingly degraded spectral resolutions.
Note that for the FUV-MAMA first-order modes, the projection of the spectrum on the detector has deliberately been shifted 120 low-resolution pixels or 3 arcseconds below center (3 arcseconds above center prior to March 15, 1999) to avoid having the spectrum fall on the shadow of the repeller wire (see also
Section 7.6 and
Section 11.1.2). This shift applies to all data taken with the
G140L and
G140M gratings, regardless of the aperture used.
The 0.2X0.2 aperture is now supported for use with all first-order gratings. This is intended to be used for observations where a long slit might allow light from another target into the aperture, thereby creating either contamination problems or bright object concerns. Note, however, that the use of such a short slit will make background subtraction more difficult, especially at wavelengths where airglow lines are important.
A number of “pseudo-aperture” positions have been defined for STIS spectroscopy which allow a target to be placed at positions other than the geometrical center of the aperture without the need to specify a POS TARG. These include the E1 and E2 positions which place the target closer to the CCD readout to minimize losses due to charge transfer inefficiency (CTI), and the D1 aperture positions, which can be used to place a faint target near the bottom of the
FUV-MAMA detector, where the dark current is significantly reduced. Note that the E1 positions may be used with any first-order STIS CCD grating. The E2 positions may only be used with
G750M and
G750L. The D1 positions may only be used with the
G140L and
G140M gratings, except for the
52X0.1D1 and
52X0.05D1 positions, which are also supported for CCD
ACQ/PEAK observations.
Here we describe these pseudo-aperture locations and their intended purposes. Note that all of these pseudo-apertures define new positions within existing apertures. As a result, the APERTURE keyword in the headers of the archived data will contain the name of the parent aperture, while the
PROPAPER keyword will contain the aperture name specified in the Phase II proposal. For example, if the Phase II proposal requests the
52X0.1D1 position, the
APERTURE keyword will be set to
52X0.1, while the
PROPAPER keyword will be
52X0.1D1.
As the STIS CCD detector has accumulated radiation damage over time, the Charge Transfer Efficiency (CTE) has decreased (see Section 7.3.7). For faint sources observed near the center of the CCD detector, this can result in loss of 18% or more of the detected signal during the readout. Since the amount of these CTE losses depends on both the observed signal and background counts, there is no simple way to correct for these losses, and they can significantly affect the shape of a measured spectrum. Noticeable effects can be seen even in well exposed spectra. In addition to its effects on the counts from the observed astronomical source, CTE effects re-distribute some of the electrons in hot pixels and cosmic rays into “tails” that lag behind during the readout. These tails add significant background noise to long exposures that is not taken into account by the
STIS Exposure Time Calculator (ETC), and which can be difficult to remove. CTE effects also result in a value for the effective dark current that is about 60% larger at the center of the detector than it is near the E1 positions (see
Figure 7.8).
All of these effects can be significantly ameliorated by moving the location of the source image on the detector closer to the amplifier, thereby reducing the number of parallel transfers that occur during the readout. To this end, so-called E1 aperture positions (52X0.05E1,
52X0.1E1,
52X0.2E1,
52X0.5E1, and
52X2E1) have been defined near row 900 on the STIS CCD detector for use with the STIS first-order gratings. The use of these aperture positions is strongly recommended for the observation of faint sources. For high signal-to-noise observations of bright targets we recommend continuing to use the regular aperture positions near the center of the detector. Extensive calibration observations were performed during Cycles 11 and 12 to ensure that the calibration at the E1 aperture positions is of the same quality as it is for sources observed at the usual location on the STIS CCD. Further information regarding the use of the E1 aperture positions can be found in
Section 7.3.8.
In 1999, the E1 aperture positions were introduced to allow first-order CCD spectra to be positioned at row 900 near the CCD readout amplifier, where the Charge Transfer Efficiency (CTE) is higher than at the standard positions. This works well; however, for G750L and
G750M spectra taken near row 900, the fringe flats have to be done using the
52X0.1 aperture rather than the
0.3X0.09 aperture, which is usually used for fringe flats near the center of the detector (see
Section 11.2.3 for a more detailed discussion of infrared (IR) fringe flats). Unfortunately, the
52X0.1 slit is shifted by about one pixel in the dispersion direction from the centers of the wider long slits. This misalignment reduces the accuracy of fringe subtraction.
To address this, we have defined three E2 aperture locations: 52X0.2E2,
52X0.5E2, and
52X2E2. When these apertures are specified, the target is placed off-center in the slit, at a position coincident with the
52X0.1E1 aperture. This improves the match between the fringes in the target and lamp spectra. Be aware, however, that the
52X0.2E2 aperture position is offset sufficiently from the physical center of the aperture that there will be noticeable changes in the aperture throughput and line-spread function.
These E2 aperture positions should only be used for
ACCUM exposures with the
G750L or
G750M gratings when fringe flats with the
52X0.1 aperture are also being done. If a peakup is desired before using the
E2 apertures, the peakup should be done using the
52X0.1E1 aperture.
The FUV-MAMA suffers from an irregular dark glow that varies unpredictably in intensity. When this glow is absent, the typical dark rate of the
FUV-MAMA detector is about 6
×10
-6 counts/pix/s. When the glow is strong, it can enhance the dark current to as much as 1
×10
-4 counts/pix/s over a large fraction of the detector. For first-order spectra, the best way to minimize this extra dark current is to put the target at a location on the detector where the extra dark current is small.
For first-order spectra of faint sources less than about 1 arcsecond in angular extent, we recommend that this be done by placing the target about 2 arcseconds above the bottom edge of the FUV-MAMA detector. Since for the
G140L and
G140M the regular aperture positions are projected about 3 arcseconds below the center (in order to avoid the shadow of the
FUV-MAMA repeller wire), an additional displacement of about -6.8
arcseconds is required in the cross-dispersion, or y, direction. This can reduce the extra dark current by up to a factor of 6 (see
Figure 4.9). For
G140L observations, the default D1 position will place the spectrum about 2 arcseconds above the bottom edge of the detector. The monthly offsetting of the spectral location (see
Figure 7.6) will shift this by as much as
±1 arcsecond. Because of the larger cross dispersion plate scale of the
G140M, variations of the default spectral position for different
G140M CENWAVE values, and the monthly spectral offsetting,
G140M spectra taken at the D1 aperture positions will be located 3 to 5 arcseconds above the bottom edge of the
FUV-MAMA detector.
The D1 apertures listed in the
Table 4.3 will be supported for first-order spectroscopic
ACCUM or
TIME-TAG observations with the
G140L and
G140M. The
52X0.1D1 and
52X0.05D1 are also supported for CCD ACQ/PEAK observations. Note that the
25MAMAD1,
F25QTZD1, and
F25SRF2D1 aperture locations are intended only for first-order
FUV-MAMA slitless spectroscopy. Users who wish to offset faint imaging targets to avoid the worst of the FUV dark current should look at
Figure 7.21 or consult with a STIS Instrument Scientist via the
Help Desk.
The throughput of the E1 and D1 apertures as a function of wavelength is similar to that of the corresponding regular positions. However, there is some vignetting of the gratings that changes the overall system throughput slightly with varying position along the slit. At the E1 positions, the overall low dispersion throughputs are decreased by 2-3%, while at the D1 position the G140L throughput is increased by 2-7%. Throughput changes for the medium resolution gratings are not well characterized, but should be similar. Since these throughput changes do not depend simply on the wavelength, but also on the grating and the position on the detector, they are handled in the pipeline calibration by the use of low-order flat fields (lfl files) rather than by a change in aperture throughput curve.
The throughput of the 52X2E2 and
52X0.5E2 positions are similar to that of the corresponding E1 positions. For the
52X0.2E2 aperture, the throughput is about 20% lower than for the
52X0.2E1 position.