0.1X0.03 Aperture


Description

This slit provides the highest spectral resolution of any STIS slit and is restricted to use with the MAMA echelle gratings. Tests during ground calibration with this slit to obtain MAMA exposures of a mono-isotopic platinum lamp source produced resolving powers in excess of 200,000.

Special Considerations

Observations with this slit require a prior PEAKUP. The peakup is performed in two stages - first with the 0.2X0.09 aperture (or one of the small neutral density apertures), and then with the 0.1X0.03 aperture.


Table 13.28: 0.1X0.03 Throughputs
% Throughput
1200
13.4
1800
31.1
2400
35.7
3000
36.3
3600
34.8
4200
32.8
4800
30.9
5400
28.9
6000
27.0

Aperture
Projected Length (arcsec)
Width (arcsec)
0.1X0.03
0.1
0.025

Figure 13.75: 0.1X0.03 Aperture Throughput as a Function of Wavelength


 