STIS CCD Detector - Living with Radiation Damage
CCD Charge Transfer Efficiency Changes
The STIS CCD is a 1024 X 1024 pixel, thinned, backside
illuminated SITe CCD, with sensitivity from below 2000 up to
about 11000. In a space environment, CCD detectors accumulate
radiation damage, which over time leads to increasing dark
current, more hot pixels, and reduced charge transfer
efficiency. After twelve years on orbit, the effects of such
radiation damage on the STIS CCD have become substantial.
The Charge Transfer Efficiency (CTE) measures the fraction of
the charge in a given pixel that is transferred to the next row
during the readout. The CTE is in practice always less than
unity because some of the charge is caught in "Traps" in the
pixel and is left behind during transfer of that pixel to the
readout amplifier. The fraction of charge lost depends on the signal levels of both the source and the background. Losses are most substantial for faint sources on low background levels. Since in a CCD, charge
must be transferred many times before reaching the
readout register, even a small decrease in CTE can have a large
effect on the measured count rate near the center of the
detector. For n transfers, the fraction of detected charge will
be CTEn. In Figure 1, we show how the measured Charge Transfer
Inefficiency (CTI=1 - CTE) of STIS has changed over the years.
For very faint sources near the center of the detector, a substantial fraction of the counts collected by the detector can easily be
lost during the readout.
Figure 1 - Measured CTI values
over the history of STIS are shown. Each color gives the results
for a different source count level. Source counts are in units
of electrons per pixel along the dispersion direction, but
integrated over a box perpendicular to the dispersion.
Most of the charge lost during a transfer will reappear during a later
transfer. This leads to the appearance of "charge tails" below the sources that can by themselves
be a considerable source of noise.
The best way to ameliorate this problem is to move the source
closer to the readout of the detector. To this end, a number of
years ago, new "E1" aperture positions were defined in the long
STIS slits to place the target near row 900 of the CCD, much
closer to the readout register. This reduces the number of
transfers during the readout by about a factor of four. In
addition to reducing CTI losses, placing the source spectrum
closer to the readout register substantially reduces
contamination by the CTE "Tails" of cosmic rays and hot pixels.
Also, the dark current is noticeably lower near the top of the
detector.
While the CALSTIS software does include an algorithm to correct
the extracted flux of a point source spectrum for CTE effects,
it cannot restore the lost signal-to-noise. Use of the E1
aperture positions is strongly recommended for all but the very
brightest or most extended spectroscopic point sources. Note
also that, while the STIS Spectroscopic Exposure Time Calculator
does not currently include correction for CTE effects, the
effects can be estimated using an iraf script available at
http://www.stsci.edu/hst/stis/software/scripts/cteloss_descrip.html
Additional information about the STIS CCD CTE and the algorithm
for correcting the fluxes of point-source spectra can be found
in Goudfrooij et al. (2006 PASP, 118, 1455).
CCD Dark Current
The CCD dark current has also continued to increase over time as
radiation damage has accumulated. Since the STIS CCD lacks a
working temperature controller or sensor, the dark current also
fluctuates with temperature (the CCD housing temperature is used
as a surrogate for the unavailable CCD chip temperature).
Figure 2 shows the dark current scaled to a housing temperature
of 18 C as a function of time at both the center of the CCD and
near row 900 at the E1 aperture position.
Figure 2 - The measured median
dark current in units of e-/pixel/s is shown as a function of
time for a position near the center of the detector (left plot) and
near the E1 aperture positions (right plot) at row 900. The red line shows
an extrapolation of the late 2004 values to the current epoch.
Time-Dependent Sensitivity Changes
All STIS modes have shown wavelength-dependent changes in
sensitivity over time. These changes are believed to result from
changes in contaminants on the optical surfaces within STIS. Initial
throughput measurements show the post-repair optical throughput are
within 2% of values expected from the extrapolation of trends
observed in 2004.
|