Background
CCD
Between launch on 14-Feb-1997 and failure of the "Side-1" electronics on
2001-May-16, CCD temperature was stabilized at -83 C. During the Side-1 era,
the CCD median dark rate rose from 0.0013 to 0.0043 e-/sec/pix. CCD temperature
cannot be stabilized with the "Side-2" electronics, so after 2001-May-16 thermal
fluctuations perturb the dark rate. It has been shown that
the temperature of the CCD housing closely follows the CCD temperature. For this reason
the CCD housing temperature (available in the telemetry value OCCDHTAV) is used by the
On the Fly Reprocessing (OTFR) to correct the dark rate. For more information please see
STIS ISR 2001-03 and
Section 7.3.1 of the STIS
Instrument Handbook. Based on an extrapolation
of previous trends, the CCD dark rate in Cycle 18 is expected to be near
0.015 e-/sec.
Because bright UV illumination temporarily increases CCD dark rate, acquisition
exposures should use the long pass filter, rather than the clear aperture.
For more on the effects of radiation damage, please see
"Living with Radiation Damage" in the Sept 2009 STAN.
MAMA Detectors
When the FUV-MAMA detector HV is initially turned on, the dark
rate can be as low as 6.4e-6 counts/pixel/s, but as the detector
operates, an extended glow centered in the upper-left hand
quadrant of the detector begins to appear. The rate at which
this glow increases has itself been increasing with time. The
unpredictable nature of this glow has made it impractical to
subtract this in the OTFR pipeline, however, mean images of this
glow have been made available to assist observers in subtracting
this from their data, and can be found at
http://www.stsci.edu/hst/stis/design/detectors/fuvDarkGlow.html.
After STIS was repaired in May 2009, the glow continued to follow pre-failure
trends. Current predictions suggest that the peak of the glow may increase by about
1e-6 counts/pixel/sec/HST-orbit.
Note that in the lower right hand corner of the detector, the glow is
usually absent, and the dark rate in that part of the detector
remains near 6.4e-6 counts/pixel/s, even after the detector HV has
been on for several orbits.
Prior to the "Side-2" failure in 2004, the
NUV-MAMA
dark rate varied between 0.0017 and 0.00067 counts/sec/pix,
depending on temperature, due mainly to phosphorescence by impurities in
the detector faceplate. After the "Side-2" repair in May 2009, the NUV-MAMA
dark rate was observed to be much higher than expected. Since May 2009, it has been
declining. As of Jan 2010, the NUV dark rate is about 0.005 counts/sec/pix. For additional
information please see
Section 7.5.2 of the STIS Instrument Handbook.
Earthshine and Zodiacal background
Sunlight reflected by the Earth or the Moon can scatter onto STIS detectors, so
the corresponding background depends on the
angle
between the target and the bright limb of the Earth or Moon.
Zodiacal background varies as a function of
helio-ecliptic
coordinates.
Both Zodiacal background and Earthshine depend on
wavelength. Tabulated background
count rates
for imaging filters and
intensities
vs. wavelength (data) are available.
Low
background scheduling can be requested in a Phase 1 proposal, and if
approved, specified in the Phase 2 proposal using the
LOW-SKY special requirement. Please see
HST Phase II Proposal Instructions for additional information.
Geocoronal
Tabulated
intensities and line widths are available for
geocoronal
emission lines at 1216, 1304, 1356, and 2471 Å.
Note that FUV-MAMA images with the clear or Ly-alpha filter will be dominated
by variable geocoronal Ly-alpha background.
Scheduling when geocoronal lines are weak can be requested in a Phase 1
proposal, and if approved, specified in the Phase 2 proposal using the
SHADOW special requirement. Please see
HST Phase II Proposal Instructions for additional information.
|