| Space Telescope Science Institute |
| WFC3 Instrument Handbook |
| help@stsci.edu |
The general operating modes of IR detectors have been described in Chapter 5. In this section we will detail the readout modes implemented in WFC3.7.7.1 Exposure Time7.7.2 MULTIACCUM ModeThe disadvantage of multiple readouts is that they are data-intensive. The HgCdTe detector array is 1024×1024 pixels, which is only about 1/16 by pixel number of the 4096×4102 UVIS array. However, since up to 16 IR readouts are used, the total data volume of a single IR exposure approaches that of a single UVIS frame. A maximum of 32 IR full array readouts can be stored in the instrument buffer, after which the content of the buffer must be dumped to the HST Solid State Recorder (SSR). A buffer dump of 16 full array reads takes about 5.8 minutes.
1. Array reset: After a fast calibration of the Analog to Digital Converters, all pixels are set to the detector bias level, with two rapid reset cycles of the entire chip.
2. Array read: The charge in each pixel is measured and stored in the on-board computer's memory. This is done as soon as practical after the second array reset in step 1. In effect, given the short delay and the time needed to read the array, a very short-exposure image is stored in memory. This is known as the zero read.
3. Multiple integration-read cycles: The detector integrates for a certain amount of time and then the charge in each pixel is read out. This step can be repeated up to a total of 15 times following the zero read during the exposure. All frames are individually stored in the on-board computer memory. Note that it takes a finite time (2.93 sec) to read the full array, so there is a time delay between reading the first and last pixel. Because this delay is constant for each read, it cancels out in difference images.
4. Return to idle mode: The detector returns to idle mode, where it is continuously flushed in order to prevent charge build-up and to limit the formation of residual images.There are 11 pre-defined sample sequences, optimized to cover a wide range of observing situations, available for the full-frame IR apertures. (See Section 7.7.4 for a discussion of the sample sequences available for the IR subarray apertures. The same names are used for the sample sequences, but the times are different.) The maximum number of reads (following the zero read) during an exposure is 15, which are collected as the signal ramps up. It is possible to select less than 15 reads, thus cutting short the ramp and reducing the total exposure time. However, the timing of the individual reads within any of the 11 sequences cannot be adjusted by the user. This approach has been adopted because optimal calibration of IR detectors requires a dedicated set of reference files (e.g., dark frames) for each timing pattern.
• One RAPID sequence: the detector is sampled with the shortest possible time interval between reads.
• Five linear (SPARS) sequences: the detector is sampled with uniform time intervals between reads, a so-called “linear sample up the ramp.” (“SPARS” is a contraction of the word “sparse.”)
• Five rapid-log-linear (STEP) sequences: the detector is initially sampled with the shortest possible time interval between reads, then uses logarithmically spaced reads to transition to a sequence of uniform samples.All 11 of the sequences above refer to readouts of the full 1024×1024 detector array. See Section 7.7.4 below for the timing sequences available for subarrays. Details of the sequences are in the following paragraphs. The timings of the individual reads are given in Table 7.8.The RAPID sequence provides linear sampling at the fastest possible speed. For the full array, this means one frame every 2.9 s, and the entire set of 16 reads completed in less than 44 s. The RAPID mode is mainly intended for the brightest targets. Due to the overheads imposed by buffer dumps (see Chapter 10), observations in this mode done continuously would have low observing efficiency.Figure 7.7: Example of STEP sequence with NSAMP=4. NSAMP+1 images are stored in the observer’s FITS image.Table 7.8: Sample times of 1024×1024 MULTIACCUM readouts in seconds. The information in this table can also be found in Table 13.4 of the Phase II Proposal Instructions.
As described in Section 7.4.4, subarrays are available in order to reduce data volume and enable short exposure times, defined in sample sequences. For a given sample sequence name, the sample times are shorter for smaller subarrays. However, only certain combinations of subarrays and sample sequences are supported by STScI. Other MULTIACCUM sequences can be used in principle but are not supported, and additional calibration observations would have to be made by the observer. The supported combinations are presented in Table 7.9. The exposure times may be found in the Phase II Proposal Instructions, Chapter 13 (Wide Field Camera 3).
See Section 13.3.6 of the GO Phase II Proposal Instructions, Table 13.4, for the sample times associated with each combination of sample sequence and subarray size. Note that the sample times for a given sample sequence name are shorter for smaller subarrays.Certain combinations of IR subarrays and sample sequences give rise to images containing a sudden low-level jump in the overall background level of the image. The cause of the artifact is under investigation (see Section 7.4.4).Table 7.9: Supported subarray sample sequences.