STScI Logo

High Speed Photometer
WF/PC-1 Instrument Science Reports

WFPC Instrument Science Report 92-11:

The Stability of Measles: an Autocorrelation Analysis

By W.B. Sparks, C. Ritchie, J. MacKenty, S. Baggett
December, 1992
Abstract excerpted from the report:

The amplitude and character of "measles" on the WF/PC-1 have been monitored by studying auto correlation functions for small sub-images. The measles appear to be stable in time after the last decontamination, although with significant changes across decontamination events. The character of the measle profile, although like a diffraction pattern in appearance, does not show the expected wavelength dependence of the far-field diffraction model.