Notice to COS Users about FP-POS
An error has been found in the COS Instrument Handbook about the direction of
spectral shifts when the FP-POS Optional Parameter is employed. This error
should only concern observers who devise their own FP-POS sequences or use
individual FP-POS values to place spectral features at specific locations on a
detector. The Handbook states that increasing FP-POS number (FP-POS=1, 2, 3, 4)
moves the spectral range on the detector to longer wavelengths. The direction
is opposite to this. An FP-POS=1 step shifts the central wavelength to a larger
value, and FP-POS=4 to a smaller one. Thus for FP-POS=AUTO, the observer will
obtain two exposures centered at longer wavelengths (FP-POS=1, 2), one at
shorter (FP-POS=4), and one at the nominal central wavelength (FP-POS=3). Note
that the overhead timing of mechanism motions as given in Table 9-4 does not
change. Increasing FP-POS position requires 3 seconds of overhead, while moving
to smaller FP-POS positions takes 70 seconds. Observers who may be affected by
the error in sign convention should contact their PCs and resubmit their Phase
II proposals.