JWST Sensitivity
During the JWST feasibility studies, we estimated sensitivities for the major
instruments using the JWST Mission Simulator (JMS).
We currently recommend the use of sensitivity estimates provided by the four instrument teams.
In general, these estimates are based upon a 10 sigma detection after a 10,000s
exposure. For both spectrographs and cameras, the teams have chosen specific
apertures rather than PSF fitting for estimating sensitivities. Hence they are
slightly conservative. Otherwise, the estimates are based upon the expected
performance (procurement specifications) for the optical components and detectors.
Note that these estimated sensitivities are generally better than the requirements
for the instruments (~15-20%) in order to provide management margin both in
verification testing and as-built performance.
In the NIR, these sensitivities assume a relatively low zodiacal background
(~ 20% higher than the lowest possible) and a low scattered light component.
The scattered light component is conservatively estimated to be:
| lambda(micron) | Scattered Light (MJy/sr) |
| 1.0 | 0.112 |
| 2.0 | 0.088 |
| 3.0 | 0.070 |
| 3.5 | 0.074 |
| 5.0 | 0.40/0.47 (NIRCam/NIRSpec) |
Instruments
Near Infrared Camera (NIRCam)
Near Infrared Spectrograph (NIRSpec)
Mid Infrared Instrument (MIRI)
Tunable Filter Imager (TFI)
NIRCam sensitivities shown here are 10-sigma detections in 10,000 seconds for
point sources extracted using an aperture with a radius of the larger of 2.5
pixels or 1.2lambda/D. The table at the end lists all values that have been plotted.
All NIRCam Sensitivities
All NIRCam filter sensitivities on one plot. Dots are wide filters, squares are
medium filters, and triangles are narrow filters. Horizontal lines indicate
filter bandpasses when they are larger than the symbols.
The next three plots show wide, medium and narrow filters separately.
Wide NIRCam Filter Sensitivities
Medium NIRCam Filter Sensitivities
Narrow NIRCam Filter Sensitivities
Table of NIRCam Sensitivities
The tables and plots show the limiting detectable continuum
(in nJy) and (narrow unresolved) line fluxes (in erg/s/cm2) as a
function of wavelength for the R=100 prism and R=1000 and R=2700
grating modes (three gratings each).
All data refer to unresolved point sources observed with a
canonical three shutter long MSA slit. Extended galaxies will experience significantly higher
slit-losses with respect to their integrated magnitudes than do point sources,
so the point source numbers should be used with caution.
The equivalent NIRSpec IFU and NIRSpec Fixed Slit sensitivities should be comparable.
In the plots, we indicate the two top-level NIRSpec sensitivity requirements:
- R=100 mode shall reach a limiting continuum flux of 132 nJy at 3.0
micron at S/N=10 in t=10.000 s.
- R=1000 mode shall reach a limiting line flux of 5.72E-19 erg/s/cm2 at
2.0 micron at S/N=10 in t=100.000 s (note the increased exposure time for
all R=1000 sensitivities).
The effective aperture size assumes that the signal is summed in the
spatial direction along the slit and across 2 pixels in the dispersion
direction. Since the resolution changes with wavelength for all modes
(and particularly the R = 100 CaF2 prism), we list
the 2 pixel spectral resolution for each mode.
The following text files provide these data in tabular form.
The MIRI science team has provided the following tables for the MIRI photometric
and spectrograph sensitivities. All estimates assume
a low zodiacal background (important below 10 microns), a point source, and
10 sigma measurement after 10,000s.
Low resolution spectroscopy (5-10 microns)
Medium resolution spectroscopy (IFU, 10,000s per setting)
| Lamda (Microns) | Line Flux (W/m^2) |
| 6.4 | 7.00E-21 |
| 9.2 | 1.00E-20 |
| 14.5 | 1.20E-20 |
| 22.5 | 6.00E-20 |
MIRI Camera
| Lambda (Microns) | microJansky |
| 5.6 | 0.2 |
| 7.7 | 0.28 |
| 10 | 0.7 |
| 11.3 | 1.7 |
| 12.8 | 1.4 |
| 15 | 1.8 |
| 18 | 4.3 |
| 21 | 8.6 |
| 25.5 | 28 |
The TFI science team has provided the following sensitivities assuming a
constant R=100 across the wavelength range.
(In practice, this will slowly oscillate from ~70 to ~150.) The extraction
aperture is 0.2" diameter at all wavelengths.
While the TFI wavelength range requirement is 1.6-2.5; 3.2-4.9 um, they provide
sensitivity estimates from 1.5-2.6 and 3.1-5.0 (the goal).
These data are also provided in tabular form.
TFI Sensitivity Table
|