WFPC2 Technical Instrument Report 02-04:
SMOV3B WFPC2 Photometry Check
-B. Whitmore, I. Heyer
A check of the photometric throughput of the WFPC2 was performed
March 31, 2002 (program ID: 8953). The standard star GRW+70d5824 was
observed with a selection of filters and the standard star was centered
in each of the four CCDs. The data indicate that any changes in the
photometric throughputs due to SM3B are less than 1% in most of the
visible wavelength filters, and less than a few percent in the UV
filters. The distribution shows a mean value of 0.34 +/- 0.26 sigma
(where sigma is defined separately for each filter-chip combination)
for all the observations, essentially consistent with an unchanged
photometric throughput. This corresponds to about 0.4% on an absolute
scale. Hence, the response of the WFPC2 was essentially unchanged by
the servicing mission. We also find that the long-term throughput
decline is consistent with the expected CTE loss.