STIS Cycle-19 Calibration Phase-1 Last Revised: 01/08/2011 Proposal ID: 12777 Title: MAMA NUV Flats PI: Elena Mason Co-I(s): Charles R. Proffitt, Michael A. Wolfe --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of Cycle-19 Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 11 Comments on orbit estimate: 11 single-orbit visits to build S/N --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: This program is aimed at obtaining NUV-MAMA flat-field observations for the construction of pixel-to-pixel flats (p-flats) with a SNR of ~100 per pixel. The flats are obtained with the DEUTERIUM-lamp and the MR grisms G230M. The actual choice of central wavelength and slit combination depends on the observed count level within each exposure. Note that STIS NUV-MAMA flats are taken every other cycles (i.e. during odd number cycles) in order to not drain the DEUTERIUM lamp lifetime. Description: Past experience and observations have shown that ~11 visits are sufficient to build a p-flat with the required SNR~100/pix (2x2 "binning", low-res mode). However the actual instrument setup (central wavelength and slit width) and exposure time might change during the cycle in order to guarantee the needed count-level of ~250000-280000 cps/frame. Past experience and observations have also shown that NUV-MAMA p-flats are mode independent, i.e. they do not depend on the wavelength. Therefore, a high-quality p-flat constructed with the G230M suffice all NUV-MAMA spectroscopic and imaging programs. The G230M flats will be taken with the slit at 5 different offset positions in order to illuminate also those detector region which are normally shadowed by the slit bars. The first visit will be performed with G230M centered at 2338A and the slit 52x0.5 (as in program 11862) for which we estimate ~23953cps. The first visit will need an exposure time of 4360. sec. However, from the second or third visit we might move to G230M/2419 (which will deliver a factor 1.17 more counts, implying a cps<=279926.) and exposure time of 3367sec, in order to possibly save at least 2.5hr of lamp lifetime. Accuracy: 1.0% (0.5% if combined with all previous P-flats) Comments on Accuracy: Accuracy is per low-res pixel (2x2 high-res pixels) Products: reference file (P-flat) and ISR as applicable --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- # Target information (copy for each target) Target Name.............................: NONE (D2 lamp) RA......................................: N/A Dec.....................................: N/A Flux (and units)........................: 280000 counts/s Visibility window (if known or relevant): N/A Comment on choice of target: Standard lamp for flat-field exposures. # Observing plan at the Phase I level. Be as complete as reasonable. As much as # possible information should be complete enough to allow easy # submission of a Phase 2 from this form. Can either describe in sentences # or use the exposure templates below. Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: Deuterium Type of Acquisition.....................: N/A Detector observation....................: NUV-MAMA Spectral Element........................: G230M Aperture................................: 52x0.5 Central Wavelength......................: 2338 or 2419 Exposure Time...........................: 1-2*4360 + 10-9*3367s BOP Predicted Local Count Rate for MAMA.: 0.26-0.27 count/s/low-res pixel BOP Predicted Global Count Rate for MAMA: 280000 count/s ETC PID IDs for Count Rates.............: [see comments] Special Requirements....................: SLIT-STEP=-3360, -1680, 0, 1680, 3360 Comments: Count rate based on observations from program 9625. The choice of central wavelengths is based upon high lamp output, and avoidance of emission lines in the spectra. MAMA plate scale along slit is 0.029 arcsec/pixel, fiducial bars are 0.5 and 0.86 arcsec, and the slit wheel steps 35 units per pixel. So the SLIT-STEP positions will cover the center of the largest fiducial, and roughly +/- 30%, and +/- 60% of the largest bar width from the center. --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: Visits should be spread across the period. In additions, it is preferred if at least 2 weeks separate visit 1 from visit 2 thus to allow for verification of the flat count level and the following visits modification as necessary. Prerequisites: PDB update? No On-board table update? Yes # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: As the lamp illumination declines, we might be obligated to change central wavelength and/or slit width in order to compensate, and so to preserve the limited lamp lifetime. ------------------------------------------------------------------------- # Link to Cycle 19 Science Fraction of science programs supported by this calibration: 47% of STIS total exposure time. List categories of science which can't be executed until after this proposal: None. List categories of science which are best performed contemporaneous with this proposal: None. --------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 4 Required turnaround on analysis (weeks): 16 Data volume (Mb needed for analysis): 100 Special software needed for analysis? (Describe): Description of analysis Plans: (1-2 paragraphs) Global count rates and engineering data will be examined to determine the extent of lamp degradation, and whether it matches predictions from current on-orbit experience. Flats will be combined, and a solution for the slit illumination and spectral distribution of the lamp will be derived and removed to create P-flats.