STIS Cycle-20 Calibration Phase-1 Template Last Revised: 31/08/2012 Proposal ID: 13135 Title: STIS CCD Spectroscopic Flats PI: Elena Mason Co-I(s): Charles Proffitt --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-20 Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 19 Comments on orbit estimate: _ 9 orbits for G430M/5216/50CCD spread across the cycle _ 10 orbits for G430M/512/52x2; --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: Obtain MR grating flats to determine the pixel-to-pixel variation for spectroscopic observations and produce the cycle 20 reference p-flat (m and l modes). The program is identical to MR grating visits of cycle 18 program 12766 executed during cycle 19. With respect to the previous cycle 18 orbits using LR gratins have been dropped: the time variation of the dust motes are expected to be sufficiently slow that an yearly monitoring is not necessary. Description: We will use the tungsten lamp and the MR grating G430M to determine the pixel-to-pixel variation of the STIS CCD in spectroscopic mode. The flat exposures will be taken with the apertures 50CCD and 52x2 at 5 offset positions in order to map -with a sufficient SNR- the whole sensitive area of the detector. The expected signal is >=1.4E6 ADU/pixels; while the expected accuracy will be <=1.5%. Accuracy: <= 1.5% Comments on Accuracy: How well the spectroscopic science data can be flattened. Products: Reference files and ISR as applicable. --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: # Target information (copy for each target) Target Name.............................:Tungsten Flux (and units)........................:Internal Tungsten Lamp Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: Tungsten lamp Type of Acquisition.....................: n/a Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G430M Aperture................................: 50CCD Central Wavelength......................: 5216 Exposure Time...........................: 21s, crsplit=no Number of Iterations ...................: 9 (X 9 visits) ETC PID IDs for Count Rates.............: Special Requirements or options.........: GAIN=4, Comments: One visit (9 flats) every ~40 days. Target or lamp..........................: Tungsten lamp Type of Acquisition.....................: n/a Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G430M Aperture................................: 52X2 Central Wavelength......................: 5216 Exposure Time...........................: 550s X 2, crsplit=no Number of Iterations ...................: 2 (X 10 visits; 1 slit-steps/visit and 5 slit-steps in total). ETC PID IDs for Count Rates.............: Special Requirements or options.........: GAIN=4, dither with special slit-pos pattern Comments: --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: Prerequisites: n/a PDB update? n/a On-board table update? n/a # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: ------------------------------------------------------------------------- # Link to Cycle 20 Science # Use the GO + GTO Phase 1 exposure listing and derived statistics # posted on WWW. Fraction of science programs supported by this calibration: 18% of STIS total exposure time. List categories of science which can't be executed until after this proposal: None. List categories of science which are best performed contemporaneous with this proposal: High S/N CCD spectroscopy why: For programs that need very high S/N, it would be best if the epoch of the flats spans the observations. So the flat program should be spread out over the whole cycle. --------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 2.5 Required turnaround on analysis (weeks): 2.5 Data volume (Mb needed for analysis): ~2.8G Special software needed for analysis? (Describe): existing python script (or IDL script) Description of analysis Plans: Do statistical analysis of data yearly to check for any significant changes in pixel-to-pixel flats. Co-add G430M flats after 8 weeks and then again at end of cycle to make new pfl files.