Cycle-21 Calibration Last Revised: 23 September 2013 Proposal ID: 13549 Title: STIS FUV MAMA Dark Monitor PI: Colin Cox Co-I(s): --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-21 Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 54 Comments on orbit estimate: 6-orbit visits repeated every 6 weeks --------------------------------------------------------------------------- Purpose: Monitor the behavior of the dark current in the FUV MAMA detector Provides data for dark count corrections for faint object observations Also provides a check on the health of the detector Description: Every six weeks a set of six exposures of 1380s is taken with the NUV MAMA with the shutter closed. The exposures are evenly spread over a six-hour SAA-free period. Accuracy: 1% Comments on Accuracy: Statistical error on the total number of counts in an image Products: Dark current images and temperature dependent formula for applying to an individual observation. Update to monitoring web page. --------------------------------------------------------------------------- Fraction of science programs supported by this calibration: 6% --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: (e.g. Take spectra of G191B2B at each supported central wavelength of each grating with the CCD and the 52x0.5 slit.) # Target information (copy for each target) Target Name.............................: DARK RA......................................: Dec.....................................: Flux (and units)........................: Visibility window (if known or relevant): Comment on choice of target: # Exposure list plan at the Phase I level. Be as complete as reasonable. # If possible information should be complete enough to allow easy # submission of a Phase 2 from this form. However, if you have to leave # some of the fields blank below, or want to provide the information # in another format, that is acceptable. (Make sure you do the BOP checking # though, as this is a phase-1 issue.) Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: e.g. DARK Type of Acquisition.....................: e.g. None Detector for observation................: e.g. STIS FUV MAMA Operating mode .........................: e.g. TIMETAG Spectral Element........................: e.g. None Aperture................................: e.g. FUV MAMA Central Wavelength......................: e.g. Exposure Time...........................: e.g. 1300s Number of Iterations ...................: e.g. 6 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: SAA free Comments: (e.g. if monitoring, specify frequency) Every 6 weeks --------------------------------------------------------------------------- # Special needs of this proposal? ------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 2 Required turnaround on analysis (weeks): 2 for each measurement Data volume (Mb needed for analysis): 0 Special software needed for analysis? (Describe): Python programs already written and used in cycle 20 Description of analysis Plans: Continue with plan established in cycle 20. Model dark rate as a function of time and temperature.