STIS Cycle-23 Calibration Phase-1 Last Revised: May 14, 2016 Proposal ID: TBD Title: STIS Focus Parallel Measurements PI: Charles Proffitt Co-I(s): TalaWanda Monroe, Linda Dressel --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-24; assume 52 weeks Total Prime Orbits: 1 External Parallel Orbits: 1 Internals or no-impact orbits: 0 Comments on orbit estimate: --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in STIS handbook as a guide. Purpose: To determine the focus offset between the STIS OII filter and that of WFC3/UVIS and ACS/WFC. This will allow any long term secular changes in the relative focus of the three instruments to be determined. Description: A UV bright star will be observed with STIS F28X50OII while parallel observations are done of a rich star field in 47 Tuc using WFC3 F410M and ACS F502N. The filters are chosen to allow phase retrieval to estimate the focus and other optical parameters from the PSF images. Simultaneous parallel observations allow a differential measure of the focus offset between the instruments without having to rely on the accuracy of the breathing model. This prime/parallel grouping is repeated 4 times for STIS+WFC3 using a small dither pattern. Due to buffer timing limitations, for only two of these 4 pairs can ACS full frame images also be included. This program is essentially identical to visit P2 of the Cycle 22 STIS/CAL program 14063, and to Cycle 23 program 14425, and will be the third annual repeat of this observation sequence. Note that the Cycle 23 version of this program is currently planned for 13-20 June, and analysis of that program should be available prior to finalizing the phase 2 for the new program. However, since year-to-year continuity is desired for this program, we do not expect those results to necessitate any changes to this program. Accuracy: < 1 micron of equivalent secondary despace Comments on Accuracy: Products: A measurement of the focus offset between STIS and ACS and WFC3. Possible recommendations for adjustments to the STIS corrector. --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- # Target information (copy for each target) Target Name.............................: HD 2072 RA.............................................: 00 24 14.5080 Dec...........................................: -71 58 49.09 Flux (and units)........................: U=9.47 Vegamag Visibility window (if known or relevant): ORIENT 255; 26-May-2017 to 13-Jul-2017 Comment on choice of target: The target is a UV bright star that gives good OII phase retrieval images while a number of bright stars are viewable in parallel by ACS and WFC3 UVIS. # Table of exposures (copy for each exposure): Target or lamp..........................: HD 2072 Type of Acquisition.....................: NONE Detector for observation................: STIS/CCD Operating mode .........................: ACCUM Spectral Element........................: MIRROR Aperture................................: F28X50OII Central Wavelength......................: Exposure Time...........................: 35s Number of Iterations ...................: 6 x 4 Gain ...................................: 4 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: 736232 Special Requirements or options.........: ORIENT 255 Comments: Repeat Prime/Parallel sequence 4 times in one orbit using a small dither pattern Some exposures will use SIZAXIS2=128 to reduce readout times. Target or lamp..........................: ANY Type of Acquisition.....................: NONE Detector for observation................: WFC3/UVIS Operating mode .........................: ACCUM Spectral Element........................: F410M Aperture................................: UVIS Central Wavelength......................: Exposure Time...........................: 360.0s Number of Iterations ...................: 1 x 4 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: 709909 Special Requirements or options.........: FLASH=15; CR-SPLIT=NO Comments: Rich star field in 47 Tuc; parallel with STIS observation Target or lamp..........................: ANY Type of Acquisition.....................: FLAT Detector for observation................: ACS/WFC Operating mode .........................: ACCUM Spectral Element........................: F502N Aperture................................: WFC Central Wavelength......................: Exposure Time...........................: 360 s Number of Iterations ...................: 1 x 2 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: 709912 Special Requirements or options.........: FLASH=100 Comments: Star field in 47 Tuc; parallel with STIS observation --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: Should be done annually at the same ORIENT=255 as was used for program 14425 and for visit P2 of program 14063. This orient will constrain the observations to occur between 26 May and 13 July, 2017. Prerequisites: n/a PDB update? no On-board table update? no # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? ORIENT=255 Other? N Describe (& justify) special requirements: A fixed orient needed so that in WFC3 and ACS images the same stars can be observed at the same locations yielding year-to-year comparisons of the focus that are not affected by variations over the detector FOV. The STIS target will be placed at the same detector location using a standard STIS ACQ. ------------------------------------------------------------------------- # Link to Cycle 24 Science Fraction of science programs supported by this calibration: 100% of STIS total external exposure time. List categories of science which can't be executed until after this proposal: none why: List categories of science which are best performed contemporaneous with this proposal: none why: --------------------------------------------------------------------------- Expected FTE weeks for analysis: 2 Required turnaround on analysis (weeks): 2 Data volume (Mb needed for analysis): 3GB Special software needed for analysis? (Describe): Use standard phase retrieval scripts for estimating the focus of stars in STIS OII images and ACS images. For the WFC3 images, Special Linda Dressel will use her customized software to perform phase retrieval analysis of large number of stars in the WFC3 images. Description of analysis Plans: Perform phase retrieval on bright stars in images from all three detectors, correct for the tabulated detector offsets, and compare results to focus model as well to results from the previous two iterations of this visit.