STIS Cycle-29 Calibration Phase-1 Last Revised: 05/14/2021 Proposal ID: ? Title: STIS MAMA Recovery from Anomalous Shutdown PI: T. Wheeler Co-I(s): A. Welty --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of Cycle-29 Total Prime Orbits: n/a External Parallel Orbits: n/a Internals or no-impact orbits: n/a Comments on orbit estimate: This is a contingency procedure. --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: This proposal is designed to permit a safe and orderly recovery of the STIS FUV or NUV MAMA detector after an anomalous shutdown. This is accomplished by using slower-than-normal MCP high-voltage ramp-ups and diagnostics. Anomalous shutdowns can occur because of bright object violations, which trigger the Global Hardware Monitor or the Global Software Monitor. Anomalous shutdowns can also occur because of MAMA hardware anomalies or failures. The cause of the shutdown should be thoroughly investigated and understood prior to recovery. Description: Anomalous shutdowns can occur because of bright object violations, which trigger the Global Hardware Monitor or the Global Software Monitor. Anomalous shutdowns can also occur because of MAMA hardware anomalies or failures. The cause of the shutdown should be thoroughly investigated and understood prior to recovery. Twenty-four hour wait intervals are required after each test for MCP gas desorption and data analysis. Event flags are used to prevent inadvertent MAMA usage. The recovery procedure consists of three separate tests (i.e. visits) to check the MAMAÕs health after an anomalous shutdown. Each must be successfully completed before proceeding onto the next. This proposal executes the same steps as Cycle 28, Proposal 16357. (1) Signal processing electronics check. The amplifiers threshold voltage is reduced from 0.48V to 0.28V; ORCOUNTS rates are monitored (MAMA HV is off during this procedure). (2) Slow, intermediate voltage high-voltage ramp-up. The MCP HV is slow- ramped to a voltage 300V below nominal. A dark time-tag exposure is taken during this partial ramp. A second dark accum exposure is taken where the event counter is cycled through X, Y, Z, W, VE, EV and OR. (3) Ramp-up to full operating voltage. As before, a dark time-tag exposure is taken during this ramp-up. A second dark accum exposure is taken where the event counter is cycled through X, Y, Z, W, VE, EV and OR. This is followed by and fold analysis test (See STIS ISR 98-02R). Accuracy: n/a Comments on Accuracy: n/a Products: n/a --------------------------------------------------------------------------- # Link to Cycle 29 Science # Use the GO + GTO Phase 1 exposure listing and derived statistics # posted on WWW. Fraction of science programs supported by this calibration: This is a contingency procedure only. List categories of science which can't be executed until after this proposal: n/a why: n/a List categories of science which are best performed contemporaneous with this proposal: n/a why: n/a --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: The recovery from an anomalous shutdown consists of three tests: a signal processing electronics check, high voltage ramp-up to an intermediate voltage and a high voltage ramp-up to the full operating voltage followed by a fold distribution test. During each of the two high voltage ramp-ups, time-tag exposures are taken and diagnostics are performed. Post-ramp dark accums with diagnostics follow. The final full ramp is followed by a fold distribution test. # Target information (copy for each target) Target Name.............................: n/a RA......................................: n/a Dec.....................................: n/a Flux (and units)........................: n/a Visibility window (if known or relevant): n/a Comment on choice of target: n/a # Exposure list plan at the Phase I level. Be as complete as reasonable. # If possible information should be complete enough to allow easy # submission of a Phase 2 from this form. However, if you have to leave # some of the fields blank below, or want to provide the information # in another format, that is acceptable. (Make sure you do the BOP #checking though, as this is a phase-1 issue.) Table of exposures (copy for each exposure as appropriate/desired): --------------------------------------------------------------------------- **** This list of exposures should be executed ONLY in the event of **** **** an anomalous FUV MAMA shutdown. **** --------------------------------------------------------------------------- Note (1) Bright Object Protection Considerations during. (a) This proposal uses lamp/aperture/mirror combinations which are designed to produce count rates of ~100,000-200,000 count/sec. They should not be changed without consultation with the P.I. (b) During the execution of the fold analysis some anode folds are disabled. Consequently, the OR counter does not provide a true representation of the OR count and so the Software Global Monitor (SGM) does not trigger until the enabled folds provide enough counts to the OR counter to trigger the SGM's threshold. To compensate, while the fold analysis is running the SGM threshold is reduced to 150,000 counts (130,000 counts are used due to ground system restrictions) in a 1.0 second interval, from its nominal value of 77,000 counts in a 0.1 second interval. (c) For the reasons discussed in (b) and to prevent light from entering the hub area, this test should only be run with the STIS external shutter closed. Visit 1, Exposure 1: Signal Processing Electronics Checkout - LV On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 3600s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-3 Non-Int. Switch on LV power supply. Set SGM Threshold = 77000, Integration Period = 0.1 secs. Set amplifier threshold to default (0.48V)- Collect a minimum of 30 telemetry points. Set amplifier threshold to 0.28V - Collect a minimum of 300 telemetry points. Special commanding: Instruction = EOHDTLV1_1. Visit 1, Exposure 2: Signal Processing Electronics Checkout - LV Off Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 50s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Special commanding: Instruction = ROLVTHDM1. Visit 1, Exposure 3: Signal Processing Electronics Checkout - Set Flag 2 Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 10s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Special commanding: Instruction = EOFLAG2. Visit 2, Exposure 1: Intermediate HV Checkout - Low Voltage On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 120s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-4 Non-Int. Must clear event flag 2 for the commanding to execute Switch on LV power supply. Special commanding: Instruction = EOHDTLV1_2. Visit 2 Exposure 2: Intermediate HV Checkout - Ramp HV to -1950V/-100V Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Time-Tag Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 1840s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Buffer_Time = 1800 Special Commanding Comments: Incrementally ramp up the detector up to 300 V below its nominal MCP operating voltage. A -100 volt ramp-up of the filed voltage (FV) voltage is also required for the detection of photons. Set SGM to Threshold = 100, Integration Period = 0.1 secs. Collect minimum of 5 telemetry samples at each voltage. Special commanding: Instruction = EOLVTHV1_2. Visit 2, Exposure 3: Intermediate HV Checkout - Cycle SGM Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Accum Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Set SGM to Threshold = 200, Integration Period = 0.1 secs. Collect minimum 5 samples X events. Collect minimum 5 samples Y events. Collect minimum 5 samples Z events. Collect minimum 5 samples W events. Collect minimum 5 samples VE events. Collect minimum 5 samples EV events. Collect minimum 5 samples OR events. Special commanding: Instruction = EOCYCLE1_2. Visit 2, Exposure 4: Intermediate HV Checkout - HV Off, LV Off, Set Flag 2 Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Ramp down detector. Set event flag 2 to prevent visit 3 from turning on the detector without human approval. Special commanding: Instruction = EOHVTHD1_2. Visit 3, Exposure 1: Full High Voltage Test - Low Voltage On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 120s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-7 Non-Int. Switch on LV power supply. Must clear event flag 2 for the commanding to execute. Special commanding: Instruction = EOHDTLV1_3. Visit 3, Exposure 2: Full High Voltage Test - Ramp to -2250V/-1500V Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Time-Tag Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 3350s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Buffer_Time = 1800 Special commanding Comments: Incrementally ramp up the detector up to its full operating voltage. Set SGM Threshold = 100, Integration Period = 0.1 secs. Collect minimum of 5 telemetry samples at each voltage. For requirements refer to TIR STIS 98-03. Special commanding: Instruction = EOLVTHV1_3. Visit 3, Exposure 3: Full High Voltage Test Ð Cycle SGM Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Accum Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Set SGM Threshold = 200, Integration Period = 0.1 secs. Collect minimum 5 samples X events. Collect minimum 5 samples Y events. Collect minimum 5 samples Z events. Collect minimum 5 samples W events. Collect minimum 5 samples VE events. Collect minimum 5 samples EV events. Collect minimum 5 samples OR events. Special commanding: Instruction = EOCYCLE1_3. Visit 3, Exposure 4: Full High Voltage Test - Fold Test Setup Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 240s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: Set SGM Threshold = 15000, Integration Period = 0.1 secs. Configure Mechanisms. Turn on HITM2 lamp at 10mA current. Special commanding: Instruction = EOFOLDSET1. Visit 3, Exposure 5: Full High Voltage Test - Accum Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Accum Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: NONE Exposure Time...........................: 300s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: This image is a flat field exposure with the HITM2 using special commanding so that the lamp remains on after the exposure. Visit 3, Exposure 6: Full High Voltage Test - Fold Test Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: n/a Exposure Time...........................: 2300s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: Set SGM Threshold = 130000, Integration Period = 1.0 secs. (a) Collect counter samples during flat field illumination Collect 5 samples X events Collect 5 samples Y events Collect 5 samples Z events Collect 5 samples W events Collect 5 samples VE events Collect 5 samples EV events Collect 5 samples OR events (b) Disable MAMA Folds: C2, C3, C4, C5, C6, R2, R3, R4, R5, R6 (c) Conduct fold analysis Collect 5 samples VE for following 19 combinations of MAMA folds: (1) Enabled: C2, R2; Disabled: C3, C4, C5, C6, R3, R4, R5, R6 (2) Enabled: C2, R3; Disabled: C3, C4, C5, C6, R2, R4, R5, R6 (3) Enabled: C3, R2; Disabled: C2, C4, C5, C6, R3, R4, R5, R6 (4) Enabled: C2, R4; Disabled: C3, C4, C5, C6, R2, R3, R5, R6 (5) Enabled: C3, R3; Disabled: C2, C4, C5, C6, R2, R4, R5, R6 (6) Enabled: C4, R2; Disabled: C2, C3, C5, C6, R3, R4, R5, R6 (7) Enabled: C3, R4; Disabled: C2, C4, C5, C6, R2, R3, R5, R6 (8) Enabled: C4, R3; Disabled: C2, C3, C5, C6, R2, R4, R5, R6 (9) Enabled: C3, R5; Disabled: C2, C4, C5, C6, R2, R3, R4, R6 (10) Enabled: C4, R4; Disabled: C2, C3, C5, C6, R2, R3, R5, R6 (11) Enabled: C5, R3; Disabled: C2, C3, C4, C6, R2, R4, R5, R6 (12) Enabled: C4, R5; Disabled: C2, C3, C5, C6, R2, R3, R4, R6 (13) Enabled: C5, R4; Disabled: C2, C3, C4, C6, R2, R3, R5, R6 (14) Enabled: C4, R6; Disabled: C2, C3, C5, C6, R2, R3, R4, R5 (15) Enabled: C5, R5; Disabled: C2, C3, C4, C6, R2, R3, R4, R6 (16) Enabled: C6, R4; Disabled: C2, C3, C4, C5, R2, R3, R5, R6 (17) Enabled: C5, R6; Disabled: C2, C3, C4, C6, R2, R3, R4, R5 (18) Enabled: C6, R5; Disabled: C2, C3, C4, C5, R2, R3, R4, R6 (19) Enabled: C6, R6; Disabled: C2, C3, C4, C5, R2, R3, R4, R5 (d) Enable MAMA folds C2, C3, C4, C5, C6, R2, R3, R4, R5, R6 at completion of exposure. (e) Check lamp stability by checking EV and VE Collect 5 samples events (EV) to check lamp stability Collect 5 samples Valid Events (VE) to check lamp stability (f) Turn off HITM2 lamp (g) Collect event counter data for detector dark count rate Collect 5 samples X dark events Collect 5 samples Y dark events Collect 5 samples Z dark events Collect 5 samples W dark events Collect 5 samples VE dark events Collect 5 samples EV dark events Collect 5 samples OR dark events (h) At completion of procedure reset SGM to nominal operating level Special commanding: Instruction = EOFOLDTST1 Visit 3, Exposure 7: Full High Voltage Test - FUV to LV Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M1) Operating mode..........................: Data Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: Ramp down detector. Special commanding: Instruction = EOHVTHD1_3 ------------------------------------------------------------------------- **** This list of exposures should be executed ONLY in the event of **** **** an anomalous NUV MAMA shutdown. **** ------------------------------------------------------------------------- Visit 4, Exposure 1: Signal Processing Electronics Checkout - LV On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 3600s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-3 Non-Int. Switch on LV power supply. Set SGM Threshold = 77000, Integration Period = 0.1 secs. Set amplifier threshold to default (0.48V)- Collect a minimum of 30 telemetry points. Set amplifier threshold to 0.28V- Collect a minimum of 300 telemetry points. Special commanding: Instruction = EOHDTLV2_1 Visit 4, Exposure 2: Signal Processing Electronics Checkout - LV Off Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 50s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Special commanding: Instruction = ROLVTHDM2 Visit 4, Exposure 3: Signal Processing Electronics Checkout - Set Flag 3 Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 10s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Special commanding: Instruction = EOFLAG3 Visit 5, Exposure 1: Intermediate HV Checkout - LV On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 120s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-4 Non-Int. Must clear event flag 3 for the commanding to execute. Switch on LV power supply. Special commanding: Instruction = EOHDTLV2_2. Visit 5, Exposure 2: Intermediate HV Checkout - Ramp HV to -1750V/-50V Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Time-Tag Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 2070s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Buffer_Time = 1800 Special commanding Comments: Incrementally ramp up the detector up to 300 V below its nominal MCP operating voltage. A -50 volt ramp-up of the photocathode (PC) voltage is also required for the detection of photons. Set SGM Threshold = 1000, Integration Period = 0.1 secs. Collect minimum of 4 telemetry samples at each voltage. Special commanding: Instruction = EOLVTHV2_2. Visit 5, Exposure 3: Intermediate HV Checkout - Cycle SGM Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Accum Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Set SGM Threshold = 1000, Integration Period = 0.1 secs. Collect minimum 5 samples X events. Collect minimum 5 samples Y events. Collect minimum 5 samples Z events. Collect minimum 5 samples W events. Collect minimum 5 samples VE events. Collect minimum 5 samples EV events. Collect minimum 5 samples OR events. Special commanding: Instruction = EOCYCLE2_2. Visit 5, Exposure 4: Intermediate HV Checkout - HV Off, LV Off, Set Flag 3 Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 420s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Ramp down detector. Set event flag 3 to prevent visit 4 from turning on the detector without human approval. Special commanding: Instruction = EOHVTHD2_2. Visit 6, Exposure 1: Full High Voltage Test - LV On Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 120s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: n/a, no detector HV BOP Predicted Global Count Rate for MAMA: n/a, no detector HV ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Sequence 1-7 Non-Int. Switch on LV power supply. Must clear event flag 3 for the commanding to execute. Special commanding: Instruction = EOHDTLV2_3. Visit 6, Exposure 2: Full High Voltage Test - Ramp HV to -2050V/-700V Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Time-Tag Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 3110s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Buffer_Time = 1800 Special commanding Comments: Incrementally ramp up the detector up to its full operating voltage. Set SGM Threshold = 1000, Integration Period = 0.1 secs. Collect minimum of 4 telemetry samples at each voltage. Special commanding: Instruction = EOLVTHV2_3. Visit 6, Exposure 3: Full High Voltage Test - Cycle SGM Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: FUV MAMA (M2) Operating mode..........................: Accum Spectral Element........................: DEFAULT Aperture................................: DEFAULT Central Wavelength......................: n/a Exposure Time...........................: 480s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Special monitoring per ISR 98-03 BOP Predicted Global Count Rate for MAMA: Special monitoring per ISR 98-03 ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Special commanding Comments: Set SGM Threshold = 1000, Integration Period = 0.1 secs. Collect 5 samples X dark events. Collect 5 samples Y dark events. Collect 5 samples Z dark events. Collect 5 samples W dark events Collect 5 samples VE dark events. Collect 5 samples EV dark events. Collect 5 samples OR dark events. Special commanding: Instruction = EOCYCLE2_3. Visit 6, Exposure 4: Full High Voltage Test - Fold Test Setup Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: n/a Aperture................................: n/a Central Wavelength......................: n/a Exposure Time...........................: 240s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: SEQ 1-3 NON-INT Special Commanding Comments: Sequence 1-3 Non-Int. Set SGM Threshold = 15000, Integration Period = 0.1 secs. Configure Mechanisms. Turn on HITM1 lamp at 3.8mA current. Special commanding: Instruction = EOFOLDSET2. Visit 6, Exposure 5: Full High Voltage Test - Accum Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Accum Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: NONE Exposure Time...........................: 120s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: This image is a flat field exposure with the HITM1 using special commanding so that the lamp remains on after the exposure. Visit 6, Exposure 6: Full High Voltage Test - Fold Test Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: n/a Exposure Time...........................: 2300s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: Set SGM Threshold = 130000, SGM Integration Period = 1.0 secs. (a) Collect counter samples during flat field illumination Collect 5 samples X events. Collect 5 samples Y events. Collect 5 samples Z events. Collect 5 samples W events. Collect 5 samples VE events. Collect 5 samples EV events. Collect 5 samples OR events. (b) Disable MAMA Folds: C2, C3, C4, C5, C6, R2, R3, R4, R5, R6 (c) Conduct fold analysis Collect 5 samples VE for following 19 combinations of MAMA folds: (1) Enabled: C2, R2; Disabled: C3, C4, C5, C6, R3, R4, R5, R6 (2) Enabled: C2, R3; Disabled: C3, C4, C5, C6, R2, R4, R5, R6 (3) Enabled: C3, R2; Disabled: C2, C4, C5, C6, R3, R4, R5, R6 (4) Enabled: C2, R4; Disabled: C3, C4, C5, C6, R2, R3, R5, R6 (5) Enabled: C3, R3; Disabled: C2, C4, C5, C6, R2, R4, R5, R6 (6) Enabled: C4, R2; Disabled: C2, C3, C5, C6, R3, R4, R5, R6 (7) Enabled: C3, R4; Disabled: C2, C4, C5, C6, R2, R3, R5, R6 (8) Enabled: C4, R3; Disabled: C2, C3, C5, C6, R2, R4, R5, R6 (9) Enabled: C3, R5; Disabled: C2, C4, C5, C6, R2, R3, R4, R6 (10) Enabled: C4, R4; Disabled: C2, C3, C5, C6, R2, R3, R5, R6 (11) Enabled: C5, R3; Disabled: C2, C3, C4, C6, R2, R4, R5, R6 (12) Enabled: C4, R5; Disabled: C2, C3, C5, C6, R2, R3, R4, R6 (13) Enabled: C5, R4; Disabled: C2, C3, C4, C6, R2, R3, R5, R6 (14) Enabled: C4, R6; Disabled: C2, C3, C5, C6, R2, R3, R4, R5 (15) Enabled: C5, R5; Disabled: C2, C3, C4, C6, R2, R3, R4, R6 (16) Enabled: C6, R4; Disabled: C2, C3, C4, C5, R2, R3, R5, R6 (17) Enabled: C5, R6; Disabled: C2, C3, C4, C6, R2, R3, R4, R5 (18) Enabled: C6, R5; Disabled: C2, C3, C4, C5, R2, R3, R4, R6 (19) Enabled: C6, R6; Disabled: C2, C3, C4, C5, R2, R3, R4, R5 (d) Enable MAMA folds C2, C3, C4, C5, C6, R2, R3, R4, R5, R6 at completion of exposure. (e) Check lamp stability by checking EV and VE Collect 5 samples events (EV) to check lamp stability Collect 5 samples Valid Events (VE) to check lamp stability (f) Turn off HITM1 lamp (g) Collect event counter data for detector dark count rate Collect 5 samples X dark events Collect 5 samples Y dark events Collect 5 samples Z dark events Collect 5 samples W dark events Collect 5 samples VE dark events Collect 5 samples EV dark events Collect 5 samples OR dark events (h) At completion of procedure reset SGM to nominal operating level Special commanding: Instruction = EOFOLDTST2 Visit 6, Exposure 7: Full High Voltage Test - NUV to LV Target or lamp..........................: Dark Type of Acquisition.....................: n/a Detector for observation................: NUV MAMA (M2) Operating mode..........................: Data Spectral Element........................: MIRROR Aperture................................: F25ND3 Central Wavelength......................: n/a Exposure Time...........................: 360s Number of Iterations....................: 1 BOP Predicted Local Count Rate for MAMA.: Note (1) BOP Predicted Global Count Rate for MAMA: Note (1) ETC PID IDs for Count Rates.............: Special Requirements or options.........: Special Commanding Comments: Ramp down detector. Special commanding: Instruction = EOHVTHD2_3 --------------------------------------------------------------------------- # Special needs of this proposal? This proposal is run to reactivate the FUV or NUV MAMA following an anomalous shutdown. Scheduling: As needed Prerequisites: (e.g., specific SMOV proposal; command development?) PDB update? (Specify table name if known e.g. SIAF). On-board table update? (Specify table name if known e.g. CCDBadPixelTable) # Special Requirements: Real time? Y Special commanding? Y Quick Data turnaround? Y (Engineering telemetry) Special timing required? N Other? N Describe (& justify) special requirements: The engineering telemetry for each step is required to evaluate the safety for proceeding. --------------------------------------------------------------------------- # Analysis Plan per Anomalous Shutdown Expected FTE weeks for analysis: 0.4 Required turnaround on analysis (weeks): 0.2 per step Data volume (Mb needed for analysis): n/a Special software needed for analysis? (Describe): none Description of analysis Plans: (1-2 paragraphs) Analysis is carried out by the Engineering Team within 24 hours of each test in the checkout. The engineering telemetry data is examined (voltages, currents, temperatures, relay position, status, and event counts) for agreement with predicted values from previous ground and on-orbit tests or model-predicted values. The time-tag images are examined for unusual spatial or temporal features. Dark and accum images are examined for unusual features, blemishes, and defects.