STIS Cycle-29 Calibration Phase-1 Last Revised: nov 24, 2021 Proposal ID: 16562 Title: STIS NUV-MAMA P-Flats PI: Amy Jones Co-I(s): --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of Cycle-29 Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 11 Comments on orbit estimate: 11 single-orbit visits --------------------------------------------------------------------------- Purpose: This program aims at obtaining NUV-MAMA flat-field observations to create a new p-flats with an SNR of ~100 per (low resolution) pixel. The flats are obtained with the Deuterium lamp and the grating G230M. However the exact instrument setup (slit width and central wavelength) might change during the cycle depending on the desired count level. Description: Past experience and observations have shown that 11 visits are sufficient to build a p-flat with the required SNR~100/pix (2x2 "binning", low-res mode). However the actual instrument setup (central wavelength and slit width) and exposure time might change during the cycle in order to guarantee the required count level. Past experience and observations have also shown that NUV-MAMA p-flats are mode independent, i.e. they do not depend on the wavelength. Therefore, a high-quality p-flat constructed with the G230M suffice all NUV-MAMA spectroscopic and imaging programs. The G230M flats will be taken with the slit at 5 different offset positions in order to illuminate the detector region normally shadowed by the slit bars. In Cycle 27, a setup with aperture=52X0.5 and cenwave=2417A resulted in an adequate count rate of ~1.8E5 counts/s. However, the count rates keep decreasing and will be too low for Cycle 29 and previous observations with cenwave=2579A have increased, being over the non-linearity limit. In Cycle 28 a special calibration program 16517 took multiple short exposures of the DEUTERIUM lamp at 9 different cenwaves. A setup with aperture=52x0.5" and cenwave=2499A, based on PID 16517, should yeild a count rate of ~2.2E5 counts/s. The cenwave 2499A was not included in 16517, so Visit 01 is a test exposure with aperture=52x0.2" and cenwave=2499 to verify the count rate is in the acceptable range. Accuracy: 1% Comments on Accuracy: Accuracy is per low-res pixel Products: Will combine flat exposures from this and past cycles to create a new reference p-flat. An ISR is also expected. --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- # Target information (copy for each target) Target Name.............................: NONE (D lamp) RA......................................: N/A Dec.....................................: N/A Flux (and units)........................: <=280000 count/s Visibility window (if known or relevant): N/A Comment on choice of target: Standard lamp for flat-field exposures. Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: Deuterium Lamp Type of Acquisition.....................: N/A Detector observation....................: STIS/NUV-MAMA Spectral Element........................: G230M Aperture................................: 52x0.2(v01), 52x0.5 Central Wavelength......................: 2499 Exposure Time...........................: 2878 sec, 10x3588 sec = 646 min BOP Predicted Local Count Rate for MAMA.: 0.2 count/s/low-res pixel BOP Predicted Global Count Rate for MAMA: ~230000 count/s for 52x0.5 ETC PID IDs for Count Rates.............: N/A Special Requirements....................: SLIT-STEP= 0, 3360, 1680, -1680, -3360 Comments: Count rate based on observations from program 16517. Count rates will change if/when instrument setups are modified. --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: About 1 visit per month, to allow inspection of the frames and timely change of setup as necessary and to avoid waste of orbits in case of lamp failure. Prerequisites: (e.g., specific SMOV proposal; command development?) PDB update? No On-board table update? No # Special Requirements: Real time? N Special commanding? Y Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: As the lamp illumination declines, we are obligated to change the central wavelength and/or slit width in order to compensate, and so to preserve the limited lamp lifetime. In case special commanding is not available for the required setting, on-board table update will be required. ------------------------------------------------------------------------- # Link to Cycle 29 Science Fraction of science programs supported by this calibration: 37% of STIS total exposure time. List categories of science which can't be executed until after this proposal: None. List categories of science which are best performed contemporaneous with this proposal: None. --------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 4 Required turnaround on analysis (weeks): 16 Data volume (Mb needed for analysis): 100 Special software needed for analysis? (Describe): existing IDL code Description of analysis Plans: (1-2 paragraphs) Global count rates and engineering data will be examined to determine the extent of lamp degradation, and whether it matches predictions from current on-orbit experience. Flats will be combined, and a solution for the slit illumination and spectral distribution of the lamp will be derived and removed to create p-flats.