STIS Cycle-32 Calibration Phase-1 Last Revised: 04/30/2024 Proposal ID: xxxxx Title: MAMA Spectroscopic Dispersion Solutions PI: D. Welty Co-I(s): M. Mingozzi --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-32; assume 52 weeks Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 7 Comments on orbit estimate: annual monitoring program --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: to monitor the accuracy of the MAMA wavelength calibration / dispersion solutions Description: Internal wavecals will be obtained at primary and secondary central wavelengths chosen to cover Cycle 32 use and to include the configurations used in this calibration program since cycle 18 (to enable long-term monitoring). This program uses the LINE lamp for most settings (for a total of approximately 1.6 hours) and the HITM2 lamp (which is brighter below 1270 A) for two of the shortest wavelength settings. All exposure times are the same as used in cycle 26, and all but one of the observations will be taken at a lamp current of 10 mA. Accuracy: 0.1-0.2 pixel Comments on Accuracy: internal wavelength precision that should be achieved in calibration exposures Products: The accuracy of the existing MAMA dispersion (_dsp) reference file will be assessed and described in a Cycle 32 calibration close-out report. If necessary, a new reference file will be created and delivered along with a supporting Instrument Science Report. --------------------------------------------------------------------------- # Link to Cycle 32 Science # Use the GO + GTO Phase 1 exposure listing and derived statistics # posted on WWW. Fraction of science programs supported by this calibration: 75.6% of STIS total exposure time (cycle 30). List categories of science which can't be executed until after this proposal: none why: List categories of science which are best performed contemporaneously with this proposal: none why: calibrations can be applied after the observations --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: Internal wavecals will be obtained for a selected number of central wavelengths of the 8 gratings (E140H, E140M, G140M, G140L, E230H, E230M, G230M, G230L) used with the MAMA detectors (FUV-MAMA,NUV-MAMA). For most exposures, TARGET=NONE and LAMP=LINE is used; LAMP=HITM2 is now used for E140H/1271 and G140M/1218. The lamp current is now 10 mA for all exposures except G230L/2376 (3.8 mA). Different apertures are used with each grating (see table below). Exposure times are the same as for cycle 26 (some increased from cycle 24). Below is a summary table with information about each exposure in this program: grating (column 1), central wavelength (column 2), aperture (column 3), and exposure time (column 4). Grat Wave Aperture Exptime (A) (s) G140L 1425 52x0.1 90. G140M 1218 52x0.1 540. G140M 1640 52x0.1 190. E140M 1425 0.2x0.06 640. E140H 1271 0.2x0.09 640. E140H 1343 0.2x0.09 590. E140H 1598 0.2x0.09 400. G230L 2376 52x0.05 48. G230M 1687 52x0.1 310. G230M 3055 52x0.1 22. E230M 1978 0.2x0.06 260. E230M 2415 0.2x0.06 135. E230M 2561 0.2x0.06 120. E230M 2707 0.2x0.06 87. E230H 1763 0.1x0.09 920. E230H 1963 0.1x0.09 930. E230H 2713 0.1x0.09 500. Comments: once per year # Exposure list plan at the Phase I level. Be as complete as reasonable. # If possible information should be complete enough to allow easy # submission of a Phase 2 from this form. However, if you have to leave # some of the fields blank below, or want to provide the information # in another format, that is acceptable. (Make sure you do the BOP checking # though, as this is a phase-1 issue.) Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52x0.1 Central Wavelength......................: 1425 Exposure Time...........................: 90. sec Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=10mA) Comments: Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G140M Aperture................................: 52x0.1 Central Wavelength......................: various (see table above) Exposure Time...........................: various (see table above) Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE or HITM2 (CURRENT=10mA) Comments: avoid strong glow orbits (if possible) for 1218 Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: E140M Aperture................................: 0.2x0.06 Central Wavelength......................: 1425 Exposure Time...........................: 640. sec Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=10mA) Comments: Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: E140H Aperture................................: 0.2x0.09 Central Wavelength......................: various (see table above) Exposure Time...........................: various (see table above) Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE or HITM2 (CURRENT=10mA) Comments: avoid strong glow orbits (if possible) for 1271 Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: NUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G230L Aperture................................: 52x0.05 Central Wavelength......................: 2376 Exposure Time...........................: 48. sec Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=3.8mA) Comments: Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: NUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G230M Aperture................................: 52x0.1 Central Wavelength......................: various (see table above) Exposure Time...........................: various (see table above) Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=10mA) Comments: Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: NUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: E230M Aperture................................: 0.2x0.06 Central Wavelength......................: various (see table above) Exposure Time...........................: various (see table above) Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=10mA) Comments: Target or lamp..........................: NONE Type of Acquisition.....................: none - internal Detector for observation................: NUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: E230H Aperture................................: 0.1x0.09 Central Wavelength......................: various (see table above) Exposure Time...........................: various (see table above) Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: (see table from IDT) BOP Predicted Global Count Rate for MAMA: (see table from IDT) ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: LAMP=LINE (CURRENT=10mA) Comments: --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: 2024 Nov-Dec Prerequisites: none PDB update? N On-board table update? N # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: avoid strong glow orbits for shortest wavelength settings, so that weak lines (from fading lamp) may be accurately measured ------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 6 weeks Required turnaround on analysis (weeks): not critical Data volume needed for analysis (Mb): 300 Mb Special software needed for analysis? (Describe): Custom (or modified) IDL software may be needed to create new dispersion (_dsp) reference files. Modifications to existing IDL software needed for analysis of echelle spectra (to make it semi-automated). Description of analysis Plans: Raw wavecal images will be self-calibrated using Calstis software. Existing wavelength solutions will be tested with available/modified software within IDL.