STIS Cycle-10 Calibration Phase-1 Last Revised: 4/11/01 Proposal ID: Title: STIS Echelle Sensitivity PI: Valenti Co-I(s): Stys - --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-10 Total Prime Orbits: 18 External Parallel Orbits: 0 Internals or no-impact orbits: 0 Comments on orbit estimate: Based on completed Phase 2 in /data/starlight2/valenti/ecal/c10/rps2/. 11 orbits on E230H, 5 orbits on E140H, and 1.3 orbits on E230M, 0.3 orbits on E140M, and a G230LB for cross-calibration. - --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: (1-2 lines) We will observe the flux standard G191B2B, obtaining echelle spectra in all primary and intermediate wavelength settings. For many of the intermediate wavelength settings, this will be the first STIS observation of a flux standard. By comparing observed and model spectra, we will construct calibration reference files describing spectroscopic sensitivity. Description: (1-2 paragraphs) We will obtain a peak S/N ratio of 30 per resolution element for E140M, E140H, and E230M, and a peak S/N ratio of 20 for E230H. The spectra will be noisier than previous flux calibration data, but all central wavelengths will be covered with sufficient S/N to obtain sensitivity curves accurate to better than 0.1% for all echelle orders. Accuracy: (A number) Peak S/N ratio of 20 per resolution element for E230H. Peak S/N ratio of 30 per resolution element for all other modes. Comments on Accuracy: (How that accuracy is to be interpreted) Tests with existing calibration spectra indicate that a fourth order polynomial is adequate to describe the ripple or blaze of echelle orders. Monte Carlo analysis indicates a maximum error of 0.1% in a fourth order polynomial fitted to data with a peak S/N ratio of 20 per resolution element. Errors of 0.1% are well below other sources of error in our flux calibration procedure. Products: (e.g. reference files, ISR, etc.) New CDBS absolute sensitivities for echelle modes. - --------------------------------------------------------------------------- # Target information (copy for each target) Target Name: G191B2B RA: 5H 5M 30.62S +/- 0.1S DEC: DEC=52D 49' 54" +/- 0.5", PLATE-ID=0005 Flux (and units): HST Standard Target Acq..............................: F28X50LP, 0.2 S Aperture................................: 0.2X0.2 Detector observation....................: FUV-MAMA, NUV-MAMA Spectral Element........................: E140M, E140H, E230M, E230H, G230LB Central Wavelength......................: All primary and intermediate Exposure Time...........................: 120-2195 sec ETC PID IDs for Count Rates.............: Not recorded Max rate & WL of hottest px.............: <1 counts/sec/pix for all modes Total rate & WL of long WL cutoff.......: <60,000 counts/sec for all modes Special Requirements....................: None. Comment: For ETC results see /data/starlight2/valenti/ecal/c10/etc.txt For existing data, see /data/starlight2/valenti/ecal/c10/existing Draft Phase 2 in /data/starlight2/valenti/ecal/c10/rps2/9999.prop - --------------------------------------------------------------------------- # Special needs of this proposal? None Scheduling: High priority. As of April 2001, 41% of archival E230M spectra and 16% of all archival echelle spectra were obtained with intermediate wavelength settings. Most of the intermediate wavelength settings have only ground calibration data based on models. This results in 15% discontinuities in flux in regions of order overlap. This is by far the largest source of error in our current flux calibration. These observations will reduce the error by a factor of 3-5, and hence should be scheduled promptly.