STIS Cycle-120 Calibration Phase-1 Template Last Revised: June 18, 2003 Proposal ID: 10040 Title: Test of New and Revised Pseudoaperture Positions PI: Proffitt Co-I(s): --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-10; assume 72 weeks) Total Prime Orbits: 2 External Parallel Orbits: 0 Internals or no-impact orbits: 0 Comments on orbit estimate: (e.g. 4 orbit visits, repeated monthly) two fringe flats at the end of the 1st visit that will be in occultation. Should this count as an internal orbit? --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: Check aperture positions and throughputs for E1, E2, D1, and WEDGEA0.6 pseudoapertures Description: Accuracy: 5% accuracy for aperture throughputs, 0.1 pixel accuracy for narrow slit positions. Comments on Accuracy: Products: Revised reference files. Documentation updates. --------------------------------------------------------------------------- # Link to Cycle 12 Science # Use the GO + GTO Phase 1 exposure listing and derived statistics # posted on WWW. Fraction of science programs supported by this calibration: 15% of STIS List categories of science which can't be executed until after this proposal: why: List categories of science which are best performed contemporaneous with this proposal: why: --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: 1st visit test new CCD pseudo apertures. Do peakup with regular central aperture, and then peakups with 52X0.05E1 and 52X0.1E1. Also take WEDGEA0.6 image, E2 spectra, and corresponding fringe flats. 2nd visit. Peakup at center of detector and then at 52X0.05D1 and 52x0.1D1. Then take G140L spectra through regular 52X2 followed by G140L with each D1 aperture. # Target information (copy for each target) Target Name.............................: BD+75D325 RA......................................: 8H 10M 49.47S Dec.....................................: 74D 57' 57.8" Flux (and units)........................: V = 9.543+/-.04 Visibility window (if known or relevant): Comment on choice of target: Will be used to test WEDGEA0.6, E1, & E2 positions. Target Name.............................: WD2126+734 RA......................................: 21H 26M 57.75S Dec.....................................: 73D 38' 44.8" Flux (and units)........................: V = 12.78 F(1400) = 5e-13 ergs/cm2/sec/ang Visibility window (if known or relevant): Comment on choice of target: Hot white dwarf with absorption line at 1400 Angstroms. Will be used to test D1 positions. Presence of absorption line makes it easier to check the alignment of the wavelength scale. Target Name.............................: CCDFLAT RA......................................: Dec.....................................: Flux (and units)........................: LAMP Visibility window (if known or relevant): Comment on choice of target: # Exposure list plan at the Phase I level. Be as complete as reasonable. # If possible information should be complete enough to allow easy # submission of a Phase 2 from this form. However, if you have to leave # some of the fields blank below, or want to provide the information # in another format, that is acceptable. (Make sure you do the BOP checking # though, as this is a phase-1 issue.) Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: MIRROR Aperture................................: WEDGEA0.6 Central Wavelength......................: Exposure Time...........................: 1 S Number of Iterations ...................: BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: SIZEAXIS2=256 Comments: Test image at new wedge location. Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: G230LB Aperture................................: 52X0.05 Central Wavelength......................: 2375 Exposure Time...........................: 1 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Peakup at regular position. Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: G230LB Aperture................................: 52X0.05E1 Central Wavelength......................: 2375 Exposure Time...........................: 1 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Test peakups at revised E1 positions. Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: G230LB Aperture................................: 52X0.1E1 Central Wavelength......................: 2375 Exposure Time...........................: 1 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Test peakups at revised E1 position. Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G230LB Aperture................................: 52X0.2E1 Central Wavelength......................: 2375 Exposure Time...........................: 11.6 Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: pos targ 0.033, 0 Comments: Check throughput at old 52X0.2E1 aperture position Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52X2E2 Central Wavelength......................: 7751 Exposure Time...........................: 38 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: E2 flux check Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52X0.5E2 Central Wavelength......................: 7751 Exposure Time...........................: 38 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: E2 flux check Target or lamp..........................: BD+75D325 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52X0.2E2 Central Wavelength......................: 7751 Exposure Time...........................: 38 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: E2 flux check Target or lamp..........................: CCDFLAT Type of Acquisition.....................: Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52X0.1 Central Wavelength......................: 7751 Exposure Time...........................: DEF Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: CCD flat for preceeding images Target or lamp..........................: CCDFLAT Type of Acquisition.....................: Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 0.3X0.09 Central Wavelength......................: 7751 Exposure Time...........................: DEF Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: CCD flat for regular 52X2 images Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: MIRROR Aperture................................: 52X0.05 Central Wavelength......................: Exposure Time...........................: 0.3 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Peakup at detector center Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: MIRROR Aperture................................: 52X0.05D1 Central Wavelength......................: Exposure Time...........................: 0.3 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Peakup at D1 position Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: MIRROR Aperture................................: 52X0.1D1 Central Wavelength......................: Exposure Time...........................: 0.3 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Peakup at D1 position Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACQ/PEAK Spectral Element........................: MIRROR Aperture................................: 52X0.1D1 Central Wavelength......................: Exposure Time...........................: 0.3 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Peakup at D1 position Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X2 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: 2.2 cnts/s observed BOP Predicted Global Count Rate for MAMA: 12000 cnts/s observed ETC PID IDs for Count Rates.............: o6gj04020 & o8aa40010 (STIS G140L) Special Requirements or options.........: Comments: Check of target flux at regular position Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X2D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X0.5D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X0.2D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X0.1D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52X0.05D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 25MAMAD1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: F25SRF2D1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 Target or lamp..........................: WD2126+734 Type of Acquisition.....................: ACQ/PEAK Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: F25QTZD1 Central Wavelength......................: 1425 Exposure Time...........................: 120 S Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: BOP Predicted Global Count Rate for MAMA: ETC PID IDs for Count Rates.............: Special Requirements or options.........: Comments: Check of target flux at D1 --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: Early in cycle 12 Prerequisites: Ground system development must be completed PDB update? SIAF On-board table update? ApertureTable # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: ------------------------------------------------------------------------- # Analysis Plan Analysis plan: Expected FTE weeks for analysis: 0 Required turnaround on analysis (weeks): 0 Data volume (Mb needed for analysis): 0 Special software needed for analysis? (Describe): Description of analysis Plans: (1-2 paragraphs) Check peakups to verify aperture locations. Check data flux and wavelength scales and update reference files (lfl, apt, apd files) if needed to adjust flux and wavelength calibration.