STIS Cycle-17 Calibration Phase-1 Last Revised: October 22, 2009 Proposal ID: 11861 Title: MAMA FUV Flats PI: Daniel J. Lennon Co-I(s): Charles R. Proffitt, Michael A. Wolfe --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of Cycle-17 Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 11 Comments on orbit estimate: 11 single-orbit visits to build S/N --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: This program will obtain FUV-MAMA flat-field observations with the Kr lamp for the construction of pixel-to-pixel flats with a S/N of 100 per low-res pixel. Description: This program will obtain a set of FUV-MAMA flat-field observations with sufficient counts to construct pixel-to-pixel flat fields (P-flats) for all modes. Approximately 11 visits will be required to construct a P-flat with S/N = 100 per low-res pixel. Experience with pre-flight and on-orbit monitoring flats show that the flat-field characteristics are in large measure color- and mode-independent, so that high-quality P-flats constructed with the G140M settings should suffice for all FUV-MAMA spectroscopic and imaging programs. This Cycle-17 calibration program calls for obtaining flats with G140M at 5 SLIT-STEP positions to illuminate regions of the detector normally shadowed by the slit fiducial bars. We will use a cenwave of 1470 Ang and the 52X0.1 slit, which will result in approximately 284,000 count/s (ignoring the small decline from one lamp exposure to the next) and achieve a signal-to-noise of 100/pixel. Accuracy: 1.0% (0.5% if combined with all previous P-flats) Comments on Accuracy: Accuracy is per low-res pixel (2x2 high-res pixels) Products: reference file (P-flat), ISR --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- # Target information (copy for each target) Target Name.............................: NONE (Kr lamp) RA......................................: N/A Dec.....................................: N/A Flux (and units)........................: 284,000 count/s (initially) Visibility window (if known or relevant): N/A Comment on choice of target: Standard lamp for flat-field exposures. Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: Krypton Type of Acquisition.....................: N/A Detector observation....................: FUV-MAMA Spectral Element........................: G140M Aperture................................: 52x0.1 Central Wavelength......................: 1470 Exposure Time...........................: 11x4740 S = 869 M BOP Predicted Local Count Rate for MAMA.: 0.27 count/s/low-res pixel BOP Predicted Global Count Rate for MAMA: 284000 count/s ETC PID IDs for Count Rates.............: [see comments] Special Requirements....................: SLIT-STEP=-3360, -1680, 0, 1680, 3360 Comments: Count rate based on observations from program 9624. The choice of central wavelengths is based upon high lamp output, and avoidance of emission lines in the spectra. --------------------------------------------------------------------------- # Special needs of this proposal? Scheduling: Best if observations occur in the Fall of 2009, after SMOV4 is complete. Visits should be scheduled such that at least 6 hours elapse between the end of one visit and the beginning of the next, in order to let the lamps cool sufficiently. Prerequisites: (e.g., specific SMOV proposal; command development?) PDB update? No On-board table update? Yes # Special Requirements: Real time? N Special commanding? N Quick Data turnaround? N Special timing required? N Other? N Describe (& justify) special requirements: As the lamp illumination declines, we are obligated to change central wavelength and/or slit width in order to compensate, and so to preserve the limited lamp lifetime. ------------------------------------------------------------------------- # Link to Cycle 17 Science Fraction of science programs supported by this calibration: 7% (of all HST GO prime and coordinated parallel usage, excludes snapshot) List categories of science which can't be executed until after this proposal: None. List categories of science which are best performed contemporaneous with this proposal: None. --------------------------------------------------------------------------- # Analysis Plan Expected FTE weeks for analysis: 4 Required turnaround on analysis (weeks): 16 Data volume (Mb needed for analysis): 100 Special software needed for analysis? (Describe): Description of analysis Plans: (1-2 paragraphs) Global count rates and engineering data will be examined to determine the extent of lamp degradation, and whether it matches predictions from current on-orbit experience. Flats will be combined, and a solution for the slit illumination and spectral distribution of the lamp will be derived and removed to create P-flats.