COS Cycle-17 Calibration Phase-1 Template Last Revised: Date 26 Jan 2010 Proposal ID: (e.g. 9026) Title: STIS PtCr/Ne Lamp Ratios PI: I. Pascucci Co-I(s): Ch. Proffitt, C. Oliveira ?? --------------------------------------------------------------------------- # Orbit estimates are TOTAL for all of cycle-17; assume 72 weeks) Total Prime Orbits: 0 External Parallel Orbits: 0 Internals or no-impact orbits: 7 Comments on orbit estimate: Lamp monitoring program --------------------------------------------------------------------------- # This section should be written with a GO audience in mind -- use # Calibration plans in WFPC-2 handbook as a guide. Purpose: To provide improved information on the ratio of the STIS wavelength calibration lamps at all wavelengths. The LINE & HITM1 lamps have faded by a factor of several since launch, and at the shortest wavelengths the fading is enough to have significantly impacted the S/N of the wavecals. The FUV flux of the HITM2 lamp has not been checked since 1997, and so a detailed comparison of all three lamps is needed to support a proper wavelength calibration for GO proposals. Description: Take a 500 s lamp G140L, narrow (52x0.1) aperture lamp exposure with each of the HITM1, HITM2, and the LINE lamps. Also take G230LB, G430L, and G750L with the same three lamps to get a full spectrum of the lamp brightness. After evaluating the lamp ratios in the G140L setting, take a 1000 s E140H-1234 narrow (0.2x0.09) aperture lamp exposure with the three lamps. Accuracy: 5% Comments on Accuracy: measure differences in fluxes between different lamp monitoring to the accuracy above to test the fading of the lamps at this level Products: Revisions to TRANS tables for auto- and GO- wavecals --------------------------------------------------------------------------- # Link to Cycle 17 Science # Use the GO + GTO Phase 1 exposure listing and derived statistics # posted on WWW. Fraction of science programs supported by this calibration: 0% List categories of science which can't be executed until after this proposal: none why: the impact could be on future GO observations that require accurate wavelength calibration List categories of science which are best performed contemporaneous with this proposal: why: --------------------------------------------------------------------------- # THE REST OF THIS FORM IS FOR INTERNAL USE --------------------------------------------------------------------------- Observation summary: Take a 500 s lamp G140L, narrow (52x0.1) aperture lamp exposure with each of the HITM1, HITM2, and the LINE lamps. Also take G230LB, G430L, and G750L with the same three lamps to get a full spectrum of the lamp brightness. After evaluating the lamp ratios in the G140L setting, take a 1000 s E140H-1234 narrow (0.2x0.09) aperture lamp exposure with the three lamps. Summary Table Grat Wav Exptime Lamp G140L 1425 500 LINE G140L 1425 500 HITM1 G140L 1425 500 HITM2 G750L 7751 500 LINE G750L 7751 500 HITM1 G750L 7751 500 HITM2 G230LB 2375 500 LINE G230LB 2375 500 HITM1 G230LB 2375 500 HITM2 G430L 4300 500 LINE G430L 4300 500 HITM1 G430L 4300 500 HITM2 ===== NOTE: for the exposures above we will use the 52x0.1 slit After evaluating the lamp ratios in G140L do: E140H 1234 1000 LINE E140H 1234 1000 HITM1 E140H 1234 1000 HITM2 NOTE: for these exposures use the 0.2x0.09 aperture Table of exposures (copy for each exposure as appropriate/desired): Target or lamp..........................: LINE Type of Acquisition.....................: none-internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52x0.1 Central Wavelength......................: 1425 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM1 Type of Acquisition.....................: none-internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52x0.1 Central Wavelength......................: 1425 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM2 Type of Acquisition.....................: none-internal Detector for observation................: FUV-MAMA Operating mode .........................: ACCUM Spectral Element........................: G140L Aperture................................: 52x0.1 Central Wavelength......................: 1425 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: LINE Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52x0.1 Central Wavelength......................: 7751 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM1 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52x0.1 Central Wavelength......................: 7751 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM2 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G750L Aperture................................: 52x0.1 Central Wavelength......................: 7751 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: LINE Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G230LB Aperture................................: 52x0.1 Central Wavelength......................: 2375 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM1 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G230LB Aperture................................: 52x0.1 Central Wavelength......................: 2375 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM2 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G230LB Aperture................................: 52x0.1 Central Wavelength......................: 2375 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: LINE Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G430L Aperture................................: 52x0.1 Central Wavelength......................: 4300 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM1 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G430L Aperture................................: 52x0.1 Central Wavelength......................: 4300 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM2 Type of Acquisition.....................: none-internal Detector for observation................: CCD Operating mode .........................: ACCUM Spectral Element........................: G430L Aperture................................: 52x0.1 Central Wavelength......................: 4300 Exposure Time...........................: 500s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: LINE Type of Acquisition.....................: none-internal Detector for observation................: FUV_MAMA Operating mode .........................: ACCUM Spectral Element........................: E140H Aperture................................: 0.2x0.09 Central Wavelength......................: 1234 Exposure Time...........................: 1000s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM1 Type of Acquisition.....................: none-internal Detector for observation................: FUV_MAMA Operating mode .........................: ACCUM Spectral Element........................: E140H Aperture................................: 0.2x0.09 Central Wavelength......................: 1234 Exposure Time...........................: 1000s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency) Target or lamp..........................: HITM2 Type of Acquisition.....................: none-internal Detector for observation................: FUV_MAMA Operating mode .........................: ACCUM Spectral Element........................: E140H Aperture................................: 0.2x0.09 Central Wavelength......................: 1234 Exposure Time...........................: 1000s Number of Iterations ...................: 1 BOP Predicted Local Count Rate for MAMA.: n/a BOP Predicted Global Count Rate for MAMA: n/a ETC PID IDs for Count Rates.............: n/a Special Requirements or options.........: Comments: (e.g. if monitoring, specify frequency)