STScI

WFPC2 Instrument Handbook for Cycle 10

TOC PREV NEXT INDEX


Cycle 4 Calibration Plan


The primary goal of the Cycle 4 calibration plan was to provide an instrument calibration for Cycle 4 GO and GTO science programs as well as monitor internal health, photometric and optical stability of the instrument. This report briefly summarizes the WFPC2 calibration proposals as they were implemented during Cycle 4. Table 8.3 outlines the proposal contents as well as the approximate frequency of execution while the following section presents more details of the justification and intent for each proposal.
Table 8.3: Summary of Cycle 4 Calibration Plans
ID Title Frequency notes
Internals
5568 Decontaminations 1 x month internals plus decon itself
5560 Internal Monitor 1. 2 x week intflats, biases, kspots
5561 5655 Internal Monitor 2. - Flats ~2 x month darks, visflats. UVflats suspended after June 12; one set of uvflats run as a test (Oct. 1994)
5562 Internal Monitor 3. - Darks 5 per week serials=ON taken in 5561/5655
5569 5764 Internal Flats once
intflats & visflats, in variety of filters
6140 Ramps: Internals & Earthcals once internals to aid in LRF calibration
Photometry
5563 Photometric Monitor - UV Std. 2 x month GRW+70d5824, std.filters. Clocks OFF except Dec. 1994, 1994 (ON).
5565 5663 Photometric Monitor - Fields 2 x month Usually NGC5139. Clocks OFF except Dec. 1994 (ON).
5629 Photometric Monitor - Four Chip Std. 1 x week GRW+70d5824; 4 CCDs, F170W
5572 Photometric Filter Calibration
run twice during Cycle 4 GRW+70d5824 in all filters used by GO/GTOs during cycle 4; NGC5139 done in subset of broadbands.
5646 CTE Dither Test run once Omega Cen, nine 40 sec images in F555W, offset in 15" steps
5659 CTE Dither Test - Part 2
run once same as 5646 but in F439W, F814W
5564 Photometric Calibration: 4-CCD to be run once GRW+70d5824, in 4 CCDs, standard broadband filters
Earth Flats
5570 Earth flats - Large Set ~ 40 per week 200 Earthflats in each of four filters: F375N, F502N, F656N and F953N
5571 Earth flats - Small Set ~14-20 per week 20 Earthflats in each of ~20 filters
Other
5643 Partially Rotated Filters done once GRW+70d5824 & visflats, FQUVN, FQCH4N and partial rotations
5632 Preflash Test done twice Omega Cen in broadband filters; includes preflash done with intflat; replaced 5565 during Apr. 1994
5778 Lyman Alpha Throughput Check do twice target: BD+75D325
5573 Ramp Calibration

5574 Polarization Calibration done once G191B2B, BD+64D106, RMon, NGC 4147, and visflats
Placeholders/Contingency (no plans to execute, may not be run at all)
5627 Photometric Monitor 1a: Std. in UV
Not needed.
5628 Photometric Monitor 1b: Std. in Vis.
Not needed.
5645 Optimum Preflash Test
Not needed.
5648 Charge Transfer Test
Withdrawn.
5566 5567 UV Campaign Photometry UV Campaign Flats
Not needed; no UV campaigns run
5575 Dither Test
Not needed.

Internal Monitors

These proposals monitor the instrument's health throughout the cycle, including the stability of the cameras, their signal chain electronics, and the internal optical alignment of the WFPC2. The internal flat fields are intended as delta-flats, in order to monitor the stability for the main photometric filter set as well as provide data for the generation of high S/N flat fields. The darks are required not only for the generation of darks for the pipeline but also for tracking the evolution of hot pixels and mapping low-level CCD defects such as traps.

The internal visflats obtained in most of the visible WFPC2 filters will be used with the thermal vacuum test flat fields and the Earth flats to generate "superflats"; the intflats are obtained as a backup for the visflats (for example, if the cal-channel should fail).

Photometric Monitors

These proposals provide a regular monitor of the instrument's QE stability from the FUV to the near-IR, allow tracking of the UV throughput decline due to contaminant buildup, and provide observational PSFs. The standard star observations provide a baseline for the standard field photometry and allow updates to SYNPHOT. The standard field measurements will allow calibration of the filter sets' color terms, as well as enabling a mapping of the geometric distortions across the field of view. The four CCD photometry program provides the relative CCD-to-CCD sensitivities for the filter set and allows the regular PC1/WF3 standard star photometry to be applied to WF2 and WF4.

Earth flats

These are divided into two proposals: the first to obtain a large number of Earth cals (observations of the bright Earth) in just four filters and the second to obtain a small set of Earth cals in a larger number of (primarily narrowband) filters. The superflats generated from the first proposal's data map the OTA illumination pattern and are combined with the thermal vacuum test data flats (and possibly cal-channel flat fields) to provide a set of flat fields which remove both the OTA illumination pattern and the pixel-to-pixel response of the cameras. The images from the second proposal will also help determine the OTA illumination pattern, but used in conjunction with the Earth superflats (and cal-channel flats), will provide delta corrections to the superflats applicable to the narrow- and medium- band filters.



Space Telescope Science Institute
http://www.stsci.edu
Voice: (410) 338-1082
help@stsci.edu