NICMOS uses this mode of operation without any slit or aperture at the input focus so that all objects in the field of view display their spectra for true multi-object spectroscopy. The NICMOS grisms operate in the spectral range between 0.8 and 2.5 µm. The grisms reside in the filter wheel for Camera 3, therefore the spatial resolution of the spectroscopy is similar to the spatial resolution of Camera 3. The filter wheel contains three grisms, of infrared grade fused silica, which cover the entire NICMOS wavelength range with a spectral resolving power of ~200 per pixel.
Since the Grisms are located exclusively in Camera 3, they are subject to the same limitations and policies as all other NIC3 observations.
Also, so far all SMOV grism tests have been obtained with NIC3 significantly out of focus. These data suggest that the grisms will perform as expected.
The NICMOS grisms have an interference filter coated on their entrance faces to limit the bandpass of the spectrum. This is necessary to prevent overlap of orders and reduce thermal background from the telescope. Since the NICMOS grisms do not have an input slit or aperture, there is not a reduction of the background flux found in slit dispersing systems. This is not a significant problem in the shorter wavelengths, but the long wavelength grism has a high background flux.
The basic parameters of the NICMOS grisms are given in Table 5.1.
Relationship Between Wavelength and Pixel
Table 5.2 gives the dispersion relationship in the form:
wavelength = m*pixel + b,
where wavelength is in microns and the 0 pixel is at the central wavelength. The relationship is plotted in Figure 5.8. The actual location of the plus and minus pixels will be dependent on the grism orientation and the location of the source in the image. The grisms are aligned as accurately as possible along a row or column of the array. We do not expect any distortion or curvature in the spectrum.
The orientation and position of the spectra relative to the direct object has been measured in orbit and has been confirmed to be identical to the Thermal Vacuum measurements. The current best estimates of the dispersion relations are those measured before launch.
Grism |
m |
b |
|---|---|---|
|
1
|
-0.005126312
|
0.9638530
|
|
2
|
-0.007379983
|
1.4098832
|
|
3
|
-0.010506387
|
2.0583025
|
Figure 5.8: Wavelength Versus Pixel Number for each Grism. Note that the actual location of the central wavelength on the detector depends on the position of the source.

Multi-Object Spectroscopy
Grism observations are carried out in the same manner as any of the imaging operations discussed earlier. In multi-object spectroscopy one of the grisms in the filter wheel for NIC3 will be selected. The observations then proceed via one of the readout and operation modes discussed later. Grism Decision Chart
The decision chart given in Figure 5.9 leads you through the construction of a grism observation.
Figure 5.9: Grism Decision Chart

Sensitivity
Background radiation will be a greater concern for grisms than for imaging observations. Every pixel on the array will receive background radiation over the spectral bandpass of the particular grism, while the source spectrum will be dispersed over many pixels. Therefore, the ratio of the source to background flux will be much lower for the grisms than for the regular imaging mode filters. The expected detected background rate per pixel is shown in Table 5.3 below for the three grisms. The increase in the background flux for grism C is dramatic. Use grisms A and B when possible. Grism C is for the longer wavelengths only.
Table 5.3: Grism Background Radiation
Grism |
Wavelength range microns |
Background(e-/sec/pixel) |
Background (Jansky/pix) |
|---|---|---|---|
|
A
|
0.8-1.2
|
0.42
|
3.2x10-5
|
|
B
|
1.1-1.9
|
1.6
|
9.5x10-5
|
|
C
|
1.4-2.5
|
360
|
0.013
|
The ETC also provides a line correction factor curve. To use this, find the wavelength of your line and read off the correction factor
G from the graph. As described in Chapter 4 multiply the line flux by this factor and add to the continuum flux. The integration time may now be calculated from the sensitivity curve as if you had a pure continuum source. To use the sensitivity curves look up the integration time required for your source flux on the sensitivity curve for the signal to noise you want. Then go to the associated exclusion zone curve and check that you are not in the shaded areas. If you are, adjust your integration time appropriately until you are in the clear area. If you are to the left of the vertical dashed line then you must use bright object mode. Grism A: G096
Figure 5.11: Grism B Throughput.
Grism C: G206
High thermal background. Use only for bright sources.
Figure 5.12: The Grism C Throughput.
Grism Analysis Software
Software is being developed at the ST-ECF for the analysis of grism observations. Using this, the observer will be able to fully extract spectra of single objects from the images, including the disentanglement of overlapping spectra and extended sources. To obtain the best results it is recommended in fields with multiple or extended sources that grism images be obtained at more than one spacecraft roll angle (preferably 3 or more), and it is essential that the image spectrum pair be obtained as described. If the matching image is not obtained, reduction and analysis of the grism data may be very difficult.
For grism spectroscopy of extended objects or crowded fields, observations of the same field with a variety of roll angles is advisable. Such data might be handled by extracting the spectra of a particular object from the image where there is the least amount of blending of the spectrum with other objects. A more promising procedure is to reconstruct a position-wavelength cube with a simultaneous deconvolution of all the grism images. Experimental software (again in IDL) has been developed at ST-ECF and further information is available at http://ecf.hq.eso.org/nicmos/nicmos.html.